ASTM D6166-12 (Reapproved 2022) is a standard test method for the quantitative, instrumental determination of the color of clear, yellow/brown liquid materials. It is specifically intended for the pine chemicals industry, covering products such as tall oil, tall oil fatty acids, rosin, and related derivatives. This method provides a more precise and objective measurement of the Gardner color scale compared to the purely visual method outlined in Test Method D1544.
The test method is applicable to materials falling within a Gardner color range of 1 to 18. It is explicitly stated that the Gardner scale is not applicable to materials lighter than 1 or darker than 18. The values stated in SI units are to be regarded as standard.
Apparatus: The test requires an instrument capable of measuring transmitted color and reporting results directly in the Gardner color scale. If such an instrument is unavailable, one may be used that reports tristimulus values or chromaticity coordinates using standard Illuminant C and the 2° observer as described in Practice E308, from which the Gardner color can be derived.
Two types of sample cells are permitted. Standard Gardner glass tubes (as per D1544) offer convenience but lower accuracy. For referee situations and where maximum precision is required, glass cuvets with a 10-mm path length must be used.
| Specification 🟦 | Standard Gardner Tubes 📏 | Glass Cuvets 📐 |
|---|---|---|
| Accuracy 🎯 | Lower accuracy | Higher accuracy (Referee use) |
| Path Length | Variable | 10 mm |
| Recommended Use | Routine checks | Dispute resolution / Certification |
Calibration and Preparation: Instruments must be calibrated and standardized following the manufacturer’s recommendations. For solid materials like most rosins and rosin derivatives, the sample must be melted and introduced into the tube or cuvet in a liquid state for measurement.
The measurement yields a color rating on the standard Gardner scale from 1 to 18. The method references Practices E691 and E177 for establishing precision and bias through interlaboratory studies. The instrumental determination provides significantly improved repeatability and reproducibility over visual comparison.
| Parameter 📊 | Specification / Standard 🎯 |
|---|---|
| Color Scale | Gardner (1 to 18) |
| Measurement Mode | Transmitted Color |
| Illuminant / Observer | Illuminant C / 2° Observer |
| Reference for Precision | Practice E691 & E177 |
| Committee Jurisdiction | D01.34 (Pine Chemicals & Hydrocarbon Resins) |
The standard is developed in accordance with the World Trade Organization (WTO) principles on international standards.
🔍 What specific products are covered under the scope of D6166-12?
This method covers pine chemical products, specifically tall oil, tall oil fatty acids, rosin, and related products.
💡 What is the valid range for Gardner color measurement according to this standard?
The test method is applicable to materials having colors from Gardner 1 to Gardner 18. The scale is not applicable to materials lighter than 1 or darker than 18.
⚡ What specific colorimetric conditions (observer and illuminant) must be used when converting to Gardner color from tristimulus values?
The color measuring instrument must use standard Illuminant C and the 2° standard observer, as defined in Practice E308.
📌 What is the required sample preparation for solid rosin in this test method?
Because rosins and rosin derivatives are typically solids, they must be heated to a molten state to be introduced into the sample tube or cuvet for the transmitted light measurement.