Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D2240 is the standard method for measuring the indentation hardness of rubber, thermoplastic elastomers, cellular materials, gel-like materials, and certain plastics using durometer instruments. The standard defines twelve durometer types—A, B, C, D, DO, E, M, O, OO, OOO, OOO-S, and R—each designed for a specific range of material hardness and specimen geometry. Durometer hardness is one of the most widely specified material properties in rubber and elastomer specifications because it is non-destructive, rapid, and directly correlates to compound formulation and service performance. This standard is not equivalent to ISO 868 or ASTM D1415 (International Hardness), and results from different methods cannot be compared directly.
The twelve durometer types differ in indentor geometry, spring force, and the scale range they cover. Type A is the most commonly used and is suitable for medium-hardness rubber compounds (neoprene, EPDM, nitrile). Type D is used for hard rubbers and rigid plastics. Type OO and the OOO variants are designed for very soft materials such as sponge rubber, gels, and foam. Type M (micro hardness) is specifically intended for specimens that are too small or thin for standard durometer types—it accommodates specimens with a thickness or cross-sectional diameter of 1.25 mm (0.050 in.) or greater, though specimens of lesser dimensions may be accommodated under specified conditions.
| 🟦 Durometer Type | 📐 Indentor Geometry | 📏 Applicable Hardness Range | 🔬 Typical Materials |
|---|---|---|---|
| Type A | Truncated cone, 0.79 mm tip diameter | 20–90 Shore A | Medium-hard rubber, soft plastics, elastomers |
| Type B | Pointed cone | Harder than Type A range | Medium-hard to hard rubber compounds |
| Type C | Spherical indentor | Medium range | Medium-density foam, sponge rubber |
| Type D | Pointed cone (sharper than B) | 20–90 Shore D | Hard rubber, rigid plastics, ebonite |
| Type DO | Modified indentor geometry | High hardness range | Very hard rubber, dense elastomers |
| Type M (Micro) | Reduced-geometry indentor (scaled-down Type A) | Full range for small specimens | O-rings, tubing, specimens ≥1.25 mm thick |
| Type OO | Spherical indentor, low-force spring | Very soft range | Soft sponge, gel-like materials, soft elastomers |
| Type OOO | Spherical, very low force spring | Ultra-soft range | Gel insoles, very soft silicone gel |
| Type OOO-S | Spherical, lowest force spring | Extremely soft range | Ultra-soft gels, biological tissue analogs |
| Type R | Spherical indentor, wide-range spring | Wide hardness range | General-purpose screening across hardness ranges |
The durometer is pressed perpendicularly onto the specimen surface with sufficient force to make firm contact. Hardness can be read as either the initial (instantaneous) indentation value or the value after a specified dwell time (commonly 15 seconds). The standard allows both approaches, but the dwell time must be reported if used. The specimen must rest on a rigid, flat, horizontal surface. For curved surfaces such as O-rings or tubing, the specimen must be supported so that the indentor contacts a locally flat area. The test is typically performed at 23 ± 2 degrees C after conditioning per ASTM D1349.
At least three measurements should be taken at different points on the specimen, each spaced at least 6 mm apart and at least 12 mm from any edge. The results are averaged and reported to the nearest whole number on the durometer scale. When using a durometer with a maximum-reading indicator (peak hold), be aware that the maximum reading may differ from the instantaneous reading—the indicator captures the highest point reached, which may be lower than the initial peak if the durometer bounces. The use of maximum-reading indicators must be identified in the test report.
Durometer hardness is sensitive to several variables that must be controlled for reproducible results. Specimen thickness is critical: if the specimen is too thin, the indentor will contact the support surface (bottoming-out), producing erroneously high readings. For Type A durometers, a minimum thickness of 6 mm is recommended; for Type M, 1.25 mm suffices. Surface condition matters—molded skin surfaces will read differently than freshly cut or buffed surfaces due to differences in surface hardness and residual mold-release agents. Temperature affects elastomer hardness because the polymer’s modulus is temperature-dependent; readings taken at elevated temperatures will be lower due to thermal softening.