Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D1829‑90 (Reapproved 1999) specifies two distinct procedures for determining the insulation resistance and volume resistivity of ceramic insulating materials at elevated temperatures ranging from 100 to 500°C. These measurements are conducted in accordance with the foundational Test Methods D 257 but rely on the special high‑temperature apparatus and handling procedures defined herein to ensure accuracy at elevated conditions.
Insulation resistance and volume resistivity data obtained through this method are critical for detecting contamination in ceramic parts and materials, as well as for providing essential design data for high‑temperature insulating devices used in demanding thermal environments.
The required apparatus includes a resistance‑measurement device with suitable sensitivity and accuracy per D 257, coupled with a specialized heating chamber. The furnace must provide uniform heat with minimal fluctuation and must include an adequate muffle (e.g., aluminum oxide) to shield the specimen from direct radiation. A grounded metallic shield, preferably of silver or stainless steel, must be provided within the furnace to electrically isolate the specimen test circuit from the heating elements.
The specimen holder is a critical assembly. The specimen is mounted under sufficient pressure between two electrode backing plates made of silver or an equivalent high‑thermal‑conductivity metal. These plates must be at least 3/4 in. (20 mm) thick to prevent warping and to ensure heat equalization. One backing plate shall be movable to allow easy insertion and removal of the specimen. The high voltage lead must be brought through the oven wall using a bushing or insulated wire with very high insulation resistance at the test temperature.
| 🟦 Apparatus Parameter | 📐 Specification Requirement |
|---|---|
| Test Temperature Range | 100 to 500°C |
| Electrode Backing Plate Material | Silver, stainless steel, or equivalent |
| Backing Plate Thickness | 3/4 in. (20 mm) minimum |
| Furnace Shield Material | Grounded metallic shield (silver or stainless steel) |
| High Voltage Lead Insulation (≤250°C) | Polytetrafluoroethylene (PTFE) or high alumina ceramic |
The standard defines two procedures tailored to different testing needs. Procedure A is intended for obtaining a curve of the resistance versus temperature characteristics of a single specimen over a wide temperature range. Procedure B is optimized for the rapid testing of large numbers of specimens at a fixed temperature.
Both procedures rely on the fundamental measurement principles of Test Methods D 257 but apply the specialized high‑temperature specimen holder and furnace configuration of this standard to guarantee accuracy and repeatability under thermal stress.
| 🎯 Feature | ⚙️ Procedure A | ⚡ Procedure B |
|---|---|---|
| Primary Application | Resistance vs. Temperature curve | Rapid fixed‑temperature testing |
| Specimen Scope | Single specimen, wide temperature range | Large number of specimens |
| Test Focus | Behavior characterization over temperature | Quality control at a specific temperature |
This test method explicitly covers the determination of insulation resistance and volume resistivity for ceramic materials at elevated temperatures between 100 and 500°C.
Procedure A is designed to generate a detailed resistance versus temperature curve for a single specimen over a broad temperature range. Procedure B is intended for rapidly testing a large number of specimens at a single, fixed temperature.
The standard requires electrode backing plates to be made of silver or an equivalent high‑thermal‑conductivity metal. They must be at least 3/4 in. (20 mm) thick to prevent warping and promote thermal equalization across the specimen.
The properties measured by D1829 shall be determined in accordance with Test Methods D 257. However, D1829 provides the necessary special high‑temperature procedures, furnace configurations, and specimen holder designs to accommodate the elevated temperature requirements not fully covered by D 257 alone.