Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D1290-95 (Reapproved 2021) is the standard test method specifically designed for the determination of sediment in water-emulsion polishes using a centrifuge. The presence of sediment is widely regarded as detrimental to the physical properties, aesthetic appearance, and overall stability of polish formulations. This method provides a consistent and reliable laboratory procedure for quantifying this unwanted component and ensuring product quality.
It is important to note that the values stated in SI units within this standard are to be regarded as the standard. The method was developed in alignment with globally recognized standardization principles established by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
The centrifuge and centrifuge tubes used in this test must conform to the specifications detailed in Paragraphs 5.1 and 5.2 of ASTM D1796 (Test Method for Water and Sediment in Fuel Oils by the Centrifuge Method). The procedure begins by measuring exactly 100 mL of the water-emulsion wax into each of two centrifuge tubes. The tubes are then stoppered and carefully balanced in their holders.
The two balanced tubes are placed on opposite sides of the centrifuge head. The sample is whirled for a precise duration of 10 minutes at a rate calculated to produce a Relative Centrifugal Force (rcf) of between 500 and 700 at the tips of the whirling tubes. The relationship between the diameter of the centrifuge swing, the rcf, and the required RPM is defined in Table 2 of Test Method D1796.
Upon completion of the 10-minute whirling period, the volume of sediment in each tube is read immediately. The reading must be estimated to the nearest 0.1 mL if the sediment interface is not perfectly distinct. The final result is expressed as a percentage of sediment, calculated by averaging the two tube readings.
| 🟦 Parameter | 📏 Specification | 📐 Key Notes / D1796 Reference |
|---|---|---|
| Sample Volume (per tube) | 100 mL | Two tubes are used for duplicate determination |
| Centrifugation Time | 10 minutes | Strict timing from the start of whirling |
| Relative Centrifugal Force (rcf) | 500 to 700 | Calculated from the swing diameter (D1796 Table 2) |
| Sediment Reading Limit | Nearest 0.1 mL | Estimation required if interface is indistinct |
| Result Reporting | Average of two readings (%) | Recorded as sediment by the centrifuge method |
| Duplicate Precision Limit | Difference ≤ 0.2 mL | Intra and interlaboratory agreement threshold |
The standard explicitly states that it has no bias, as the results are defined strictly by the method itself. For precision, interlaboratory and intralaboratory duplicate determinations of sediment by this test method should not differ by more than 0.2 mL.
🔍 Why is sediment testing crucial for water-emulsion polishes?
Sediment is generally believed to negatively affect key polish properties, including gloss, durability, and application smoothness. This centrifuge method provides a standardized, objective measure of sediment content for quality assurance and formulation control.
💡 How do I determine the correct RPM for the centrifuge?
The correct RPM is not a fixed value. It must be derived from the specific diameter of swing of your centrifuge to achieve an rcf of 500–700. Table 2 of Test Method D1796 (Rotation Speeds Applicable for Centrifuges of Various Diameters of Swing) provides this necessary data for calibration.
⚡ How is the percentage of sediment calculated?
Two 100 mL samples are tested. After centrifugation, the sediment volume in each tube is read to the nearest 0.1 mL. The arithmetic mean (average) of these two readings is taken as the final percentage of sediment by the centrifuge method.
📌 What is the acceptable variation between duplicate test results?
The precision statement in Section 7.1 specifies that duplicate determinations of sediment (whether performed in the same laboratory or different laboratories) should not differ by more than 0.2 mL for the results to be considered valid.