IEC 60283: DC Resistive Volt Ratio Boxes — The Reference Standard for Precision Voltage Ratio Measurement

From Standard Cells to Quantum Voltage: How Volt Ratio Boxes Achieve 0.0001% Ratio Accuracy

IEC 60283 specifies DC resistive volt ratio box requirements. Precision ratio boxes are calibration laboratory core equipment — dividing a known standard voltage by precise ratios (10:1, 100:1, 1,000:1) for calibrating voltmeters and potentiometers.

RatioAccuracy ClassTypical Resistance (Input Z)Application
10:10.001–0.01%100kΩ/10kΩGeneral voltmeter calibration
100:10.001–0.005%1MΩ/10kΩPrecision DVM calibration
1,000:10.0005–0.002%10MΩ/10kΩStandard cell comparison, quantum voltage
Ratio accuracy:
Ratio = (R1+R2)/R2 (valid only when Z_load >> R2)
Error = TCR_mismatch × ΔT + self-heating + loading
Self-heating: P=I²R raises resistor temp → TCR drift
→ Applied voltage must be limited so P < 0.01 × P_rated

TN Lab — Precision voltage measurement begins with the accuracy of the voltage ratio.