Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
CSA C22.2 No. 60601-2-28-18 is the Canadian adoption of the international standard IEC 60601-2-28:2010, including Amendment 1:2013. This standard specifies particular requirements for the basic safety and essential performance of medical electrical equipment that is intended for the generation and application of X-rays for diagnostic purposes. It covers equipment such as X-ray generators, X-ray tubes, and integrated systems used in radiography, fluoroscopy, and computed tomography. The standard is part of the CSA C22.2 series, which harmonizes with Canadian electrical safety requirements, and it applies to both transient and permanent medical installations.
The standard establishes requirements in several critical areas to ensure safe operation and accurate diagnostic performance. Below are the major technical categories:
Manufacturers must implement measures to limit unnecessary radiation exposure. This includes design requirements for beam limitation devices, accessory performance (e.g., grids, compression paddles), and integration of exposure control systems that follow the principle of as low as reasonably achievable (ALARA). The standard mandates that X-ray equipment provide clearly labeled controls and indicators for exposure parameters such as tube voltage (kV), tube current (mA), and exposure time (s).
Given the high voltages involved (often >100 kV), the standard requires specific clearance and creepage distances, especially in high-voltage cables and connectors. It also stipulates testing for dielectric strength and partial discharge to avoid arcing that could compromise patient safety. The equipment must withstand overvoltage conditions without causing hazardous situations.
To maintain diagnostic accuracy, the standard sets acceptance limits for key parameters. These limits are verified under reference conditions (e.g., 0.5 mm Cu filtration, 1 m distance, standard detector). The table below summarizes typical performance requirements outlined in the standard:
| Parameter | Requirement (Reference Conditions) | Test Method |
|---|---|---|
| Accuracy of tube voltage (kV) | Within ±10% of indicated value or ±1 kV, whichever is greater | Non-invasive kV meter |
| Reproducibility of radiation output | Variation <10% over ten consecutive exposures at same settings | Ionization chamber |
| Linearity of dose with mAs | ±20% over the entire range | Multiple exposures at different mAs |
| Minimum alarm threshold | 10 μGy/h at 1 m from the X-ray source (leakage) | Survey meter |
Implementing CSA C22.2 No. 60601-2-28-18 requires attention to both design and documentation. Manufacturers should consider the following:
To legally sell diagnostic X-ray equipment in Canada, manufacturers must demonstrate compliance with CSA C22.2 No. 60601-2-28-18. This involves:
The standard also includes transitional provisions. As of 2026, any changes to referenced standards (like IEC 60601-1-2 for EMC) must be incorporated. Manufacturers should maintain active surveillance of amendments issued by CSA and SCC.
While CSA C22.2 No. 60601-2-28-18 is a comprehensive document, it does not cover every possible design. It is the manufacturer’s responsibility to perform a complete risk analysis and address any hazards not specifically mentioned in the standard. Always refer to the latest version of CSA C22.2 No. 60601-1 and consult with a qualified test lab.
This article is intended for informational purposes only. For official compliance, always consult the current version of CSA C22.2 No. 60601-2-28-18 and your certification body. © 2026 Technical Standards Press.