Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
CAN/CSA C22.2 No. 60601-2-68-15 is the Canadian national adoption of the international standard IEC 60601-2-68:2014, which specifies particular requirements for the basic safety and essential performance of X-ray equipment used for radiography and radioscopy (including fluoroscopy). This standard applies to medical electrical equipment and systems that generate X-rays for diagnostic imaging, excluding radiotherapy devices, computed tomography (CT) scanners, and dental extraoral X-ray equipment.
The standard is intended for manufacturers, testing laboratories, and healthcare facilities that design, install, or operate stationary and mobile X-ray systems used in general radiography, fluoroscopy, interventional imaging, and related procedures. Its scope covers both the X-ray source assembly (e.g., X-ray tube, collimator) and the complete system including image receptors, patient supports, and control devices.
This standard replaces earlier requirements found in CSA C22.2 No. 114 (for diagnostic X-ray equipment) and aligns with the latest international consensus on radiation safety, dose management, and performance testing. It harmonizes with other relevant collateral standards such as IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), and IEC 60601-1-8 (alarm systems).
The standard establishes performance criteria and test methods for a wide range of parameters that directly affect patient and operator safety. The following table summarizes the most critical technical requirements:
| Requirement Category | Specific Parameter | Acceptance Criterion (example) |
|---|---|---|
| Radiation output | Air kerma rate | ± 20% of indicated value |
| Beam quality | Half-value layer (HVL) | ≥ 2.3 mm Al at 70 kVp |
| Collimation | Light field/X‑field alignment | ≤ 2% of SID |
| Patient dose | Dose area product (DAP) accuracy | ± 25% of reference |
| Image quality | Limiting spatial resolution | ≥ 2.5 lp/mm (flat panel) |
| Electrical safety | Leakage current (patient applied parts) | ≤ 10 µA under normal conditions |
| Radiation protection | X-ray tube housing leakage | < 1 mGy/h at 1 m |
In addition, the standard mandates robust dose reporting and management features. Equipment must provide real‑time display of air kerma (AK) and dose area product (DAP) for every exposure, including cumulative dose monitoring during fluoroscopic procedures. This supports the principle of As Low As Reasonably Achievable (ALARA) and facilitates compliance with provincial regulations on fluoroscopy dose tracking.
Beyond safety, the standard defines essential performance criteria that directly relate to the clinical effectiveness of the device. These include:
When installing or upgrading radiographic and fluoroscopic (R&F) systems in Canada, healthcare providers must ensure that the equipment is certified to CAN/CSA C22.2 No. 60601-2-68-15. Key implementation steps include:
Manufacturers seeking to market diagnostic X‑ray equipment in Canada must demonstrate compliance with CAN/CSA C22.2 No. 60601-2-68-15. The certification process generally involves:
A recognized certification body (CB) performs full type tests on a representative sample of the equipment. Tests cover electrical safety, radiation safety, mechanical hazards, and essential performance as per the standard. The CB will verify compliance with both the collateral standards (e.g., CAN/CSA C22.2 No. 60601‑1) and the particular standard.
ISO 13485 certification is typically required to maintain consistency of the certified design. The CB may conduct periodic factory inspections to ensure ongoing conformity.
Equipment must be permanently marked with the certification mark and with specific warnings required by the standard. The accompanying documents must include all technical data (e.g., radiation output characteristics, filtration information, and dose values) as specified in Clause 7.
Additionally, the standard has been harmonized with Canadian provincial radiation protection regulations. In most provinces, a valid certification to this standard is a prerequisite for registering the X‑ray unit with the regulatory authority (e.g., Ontario’s Healing Arts Radiation Protection Act, British Columbia’s X‑Ray Regulations).
As of 2026, the current edition remains CAN/CSA C22.2 No. 60601-2-68-15. However, manufacturers should monitor the release of the next edition (likely referencing IEC 60601‑2‑68:2023) to plan for recertification cycles. The CSA Group typically allows a transition period of 2‑3 years after publication of a new edition.
This article provides an overview of CAN/CSA C22.2 No. 60601-2-68-15 and is intended for informational purposes. For detailed compliance guidance, consult the full standard text and engage a qualified certification body.
© 2026 – All rights reserved.