Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
CAN/CSA C22.2 No. 60601-1-3-09 (R2014) is the Canadian adoption of the international collateral standard IEC 60601-1-3:2008, titled Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral standard: Radiation protection in diagnostic X-ray equipment. Developed under the Canadian Electrical Code umbrella (C22.2 series), this standard applies to medical electrical equipment intended for diagnostic X‑ray imaging, including radiographic, fluoroscopic, mammographic, and computed tomography systems. Its primary objective is to ensure that radiation protection features are integrated into the design, manufacture, and testing of such equipment, thereby safeguarding patients, operators, and the public.
The standard covers the basic safety and essential performance aspects of radiation protection that are not addressed by the general standard IEC 60601‑1 (or its Canadian equivalent). It is mandatory for manufacturers seeking compliance under Canada’s Medical Devices Regulations and is frequently referenced by provincial regulators and accreditation bodies.
The collateral standard defines a set of technical requirements organized around the following core areas:
Requirements address the inherent and total filtration of the X‑ray beam, which influence beam quality and patient dose. The standard mandates minimum equivalent filtration (typically 2.5 mm Al for equipment operating above 70 kV) and requires that the half‑value layer (HVL) meet specified minima. Leakage radiation from the X‑ray source assembly must not exceed 1.0 mGy/h at a distance of 1 m from the focal spot when the tube is operated at its maximum rated conditions.
All equipment must provide a means to limit the useful X‑ray beam to the image receptor area. For equipment that allows field size adjustment, the deviation between the indicated and actual beam dimensions is strictly controlled (e.g., ±2 % of the source‑to‑image distance for general radiography). Automatic collimation systems must ensure the beam never exceeds the receptor area by more than 3 % of the SID. These requirements are essential to avoid unnecessary irradiation of tissues outside the region of interest.
Equipment that provides air kerma or dose‑area product (DAP) indications must meet accuracy criteria defined in the standard. For systems with automatic exposure control (AEC), the reproducibility and accuracy of the indicated dose are verified under reference conditions. The standard also specifies requirements for manual exposure control, including clear labelling of control settings and the presence of an audible or visual exposure termination indicator.
| Parameter | Requirement | Reference Clause |
|---|---|---|
| Total beam filtration (equipment ≥70 kV) | ≥ 2.5 mm Al equivalent | 201.8.1 |
| Leakage radiation (max. at 1 m) | ≤ 1.0 mGy/h | 201.9.1 |
| Beam field deviation (indicated vs. actual) | ±2 % of SID | 202.3.1 |
| Dose indication accuracy (air kerma or DAP) | ±35 % for reference conditions | 203.1.2 |
| Half‑value layer at 80 kV | ≥ 2.9 mm Al (for total filtration ≥2.5 mm Al) | 201.8.2 |
Manufacturers must integrate these radiation protection requirements throughout the design‑control process. Compliance is demonstrated through type testing, design review, and routine production testing. The standard requires that a risk management process (per ISO 14971 or CAN/CSA‑ISO 14971) is applied to identify potential radiation hazards and to verify that residual risks are reduced to acceptable levels.
Testing for certification typically involves measurement of filtration (HVL and equivalent), leakage radiation with the X‑ray tube shielded, and beam alignment/field size accuracy. Dose indication accuracy is verified using calibrated reference detectors under the conditions defined in the standard. Special attention must be paid to equipment intended for mammography or CT, where additional clauses (notably 201.10 for mammography and 201.11 for CT) impose more stringent limits on dose and beam quality.
In Canada, recognition of CAN/CSA C22.2 No. 60601-1-3-09 is often a prerequisite for Health Canada medical device licensing. Most major certification bodies issue a certificate of compliance that references this standard alongside the general standard CAN/CSA C22.2 No. 60601-1-08 (adoption of IEC 60601-1). Canadian deviations include additional markings (e.g., bilingual warnings) and alignment with the Canadian Electrical Code Part I for installation requirements.
It is important to note that the standard was reaffirmed in 2014 (R2014) and remains current as of 2026. However, IEC has published a 2022 edition of IEC 60601-1-3 (Edition 2.1). Manufacturers planning new product introductions should monitor the CSA adoption schedule for potential updates that may introduce stricter requirements (e.g., for pediatric patients and interventional fluoroscopy).