📐 PD ISO/TS 17379 is a BSI Published Document covering the determination of selenium in water. Part 1 uses hydride generation atomic fluorescence spectrometry (HG-AFS) and Part 2 uses hydride generation atomic absorption spectrometry (HG-AAS). Selenium is a trace element of major environmental and health interest: at low levels it is an essential nutrient, but at higher levels it is toxic. Accurate determination of selenium in drinking water, surface water, groundwater and industrial wastewater is therefore essential for water monitoring, environmental assessment and drinking-water safety. The series was prepared by BSI Technical Committee EH/3/2 (physical chemical and biochemical methods) and published in July 2013.
PD ISO/TS 17379 applies to the determination of selenium in water, particularly at trace levels. Part 1 (HG-AFS) specifies the principle, interferences, reagents and standard solutions, apparatus, sample preparation, procedure, calculation and precision for the determination of selenium by hydride generation atomic fluorescence spectrometry. Part 2 (HG-AAS) specifies the corresponding content for the determination of selenium by hydride generation atomic absorption spectrometry. Part 2 supersedes the withdrawn BS 6068-2.45:1993.
| Part | Reference | Detection Principle | Typical Application |
|---|---|---|---|
| Part 1 | PD ISO/TS 17379-1:2013 | Hydride generation atomic fluorescence spectrometry (HG-AFS) | Trace selenium, low-matrix waters, high sensitivity |
| Part 2 | PD ISO/TS 17379-2:2013 | Hydride generation atomic absorption spectrometry (HG-AAS) | Routine water monitoring, medium concentrations |
The common core of both methods is hydride generation. In acid medium, selenium(IV) reacts with sodium borohydride (NaBH₄) or potassium borohydride to produce gaseous hydrogen selenide (H₂Se):
Se(IV) + NaBH₄ + H⁺ → H₂Se↑ + by-products
The H₂Se is carried by an inert gas (e.g. argon) into an atomiser. In HG-AFS, H₂Se is atomised in a hydrogen-argon flame or electrically heated quartz tube; the excited selenium atoms emit characteristic fluorescence whose intensity is proportional to selenium concentration. In HG-AAS, H₂Se is thermally decomposed to atomic selenium and quantified by measuring absorbance at a characteristic line (e.g. 196.0 nm). Both methods use standard curve quantification.
In hydride generation atomic spectrometry for selenium, interference mainly comes from coexisting metal ions and volatile hydride-forming elements. Common interfering factors include:
To remove these interferences, the standard specifies reagent purification, masking agents, acidity adjustment, carrier-gas optimisation and matrix matching, and requires spike recovery and precision to be verified during method validation.
Key reagents include: selenium stock and working standard solutions, sodium (or potassium) borohydride reductant solution, acid (e.g. hydrochloric acid), anti-interference masking agents, and matrix-matched solutions for preparing calibration standards. For apparatus, HG-AFS requires a hydride generation atomic fluorescence spectrometer, while HG-AAS requires a hydride generation atomic absorption spectrometer (or an AAS with a quartz atomiser).
A typical procedure is: sample collection and preservation → digestion/pre-reduction → acidity adjustment → hydride generation → atomisation and detection → quantification against a standard curve → calculation and reporting. The whole procedure should include reagent blanks, duplicate samples and QC samples to ensure reliability.