Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Radial lip force, also known as radial load or lip force, is a critical characteristic of radial lip seals that determines how effectively the seal contacts the shaft. SAE J1901, though cancelled, remains a definitive reference for the measurement principle and practices. This article summarizes the key aspects of lip force measurement using the split shaft method, compares measurement devices, and highlights best practices for consistent and accurate results.
Most devices for measuring radial lip force use a split shaft design. The shaft is divided into two halves with a gap of 0.75 to 1.50 mm to prevent contact between the halves. One half is rigidly mounted; the movable half is attached to a force sensing system. When the seal is installed over the shaft, it exerts a force on the movable half, which is measured as force P. The lip force per unit circumference F is then calculated as:
F = P / D and T = F × πD
where D is the shaft diameter. The measured force P is the component perpendicular to the split.
SAE J1901 describes three types of devices for converting the force from the split shaft into a measurable reading: mechanical, pneumatic, and electronic. Each has distinct characteristics that affect measurement accuracy.
| Device Type | System Stiffness | Advantages | Limitations |
|---|---|---|---|
| Mechanical (leaf spring + dial indicator) | Low – deflects significantly | Simple, low cost | Errors from effective shaft size change; sensitive to lip position and weight of large shafts |
| Pneumatic (air cylinder) | Effectively perfect – no deflection | High stiffness, no shaft deflection | Suspension errors possible; more complex setup |
| Electronic (strain gauge transducer) | High – very small deflections | Accurate, direct readout, diameter divider for immediate F value | Requires calibration; higher initial investment |
🛠️ Engineering Insight: Electronic devices are preferred because their high system stiffness minimizes measurement errors due to shaft size change. Combined with a digital readout that includes a time delay, they provide reliable and repeatable readings.
When a seal is first seated on the shaft, the lip force decays exponentially over time due to stress relaxation. The rate of decay depends on the polymer type. SAE J1901 recommends a standard time delay of 30 seconds for all materials before taking the reading, which covers even the slowest relaxing material (fluoroelastomer).
| Polymer Type | Approximate Time to Constant Decay Rate (seconds) |
|---|---|
| Nitrile | 25 |
| Polyacrylate | 10 |
| Silicone | 3 |
| Fluoroelastomer | 30 |
Seal temperature significantly affects radial force measurements. To standardize results, seals should be stored at 21 °C ± 3 °C for at least 24 hours before measurement. The temperature sensitivity differs by material, as shown below:
| Polymer Type | Approximate Change per ±11 °C from 21 °C |
|---|---|
| Nitrile | ±6% |
| Polyacrylate | ±8% |
| Silicone | ±3% |
| Fluoroelastomer | ±12% |
Dead weight calibration should be performed monthly to ensure accuracy. When making multiple measurements on the same seal, allow 8 hours between readings to let the material relax fully.
This gap is small enough that it does not affect the lip force distribution, but large enough to prevent contact between the two shaft halves, which would otherwise produce erroneous force readings.
A low‑stiffness system (e.g., mechanical leaf spring) can deflect under load, changing the effective shaft diameter and introducing measurement errors. High‑stiffness systems, like electronic transducers, minimize deflection and improve accuracy.
Stress relaxation causes the lip force to decay rapidly after the seal is placed on the shaft. A 30‑second delay ensures that the reading is taken after the initial decay, providing a stable and repeatable value. This time period covers the longest relaxation time observed across common elastomers (fluoroelastomer: 30 s).
Seals should be stored at 21 °C ± 3 °C for 24 hours prior to measurement. This temperature was chosen for convenience and standardization; deviations can cause significant errors, as shown in the temperature sensitivity table above.
This article summarizes the core principles of SAE J1901, offering engineers a practical reference for radial lip force measurement. Always refer to the latest industry standards and the seal manufacturer’s recommendations for specific applications.