Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
The accuracy of automated non-invasive blood pressure (NIBP) monitors is a cornerstone of effective cardiovascular diagnosis and patient management. Serving as the definitive global benchmark for clinical performance assessment, ISO 81060-2:2019 (published in Canada as CSA ISO 81060-2:19) outlines the rigorous protocol that manufacturers must follow to validate their devices. This article details the essential technical facets of this standard, focusing on scope, strict validation criteria, and implementation pathways.
ISO 81060-2:2019 specifies the requirements and methods for the clinical investigation of intermittent, automated, non-invasive blood pressure measurement devices. This includes devices used on the upper arm, wrist, and thigh. The standard provides an unambiguous framework for determining the systematic bias (mean error) and precision (standard deviation) of a test device when compared against a clinically accepted reference standard. It replaces the grading system of the 2013 edition with a definitive binary pass/fail system, significantly tightening the pathway to market access.
The standard is highly prescriptive in its methodology to minimize bias and ensure statistical power across diverse patient populations. It mandates a sequential same-arm comparison between the test device and the reference standard.
A critical element of the protocol is the required distribution of blood pressure among the subjects. This ensures the device is tested across the full physiological range of intended use. The table below summarizes the required distribution for Systolic Blood Pressure (SBP).
| SBP Baseline Range (mmHg) | Required Proportion of Subjects |
|---|---|
| < 100 | 5% – 15% |
| 100 – 129 | 20% – 40% |
| 130 – 159 | 25% – 50% |
| 160 – 179 | 10% – 30% |
| ≥ 180 | 5% – 15% |
The 2019 edition introduces a rigorous binary pass/fail system. To pass, the device must meet two criteria for both SBP and Diastolic Blood Pressure (DBP):
A sample size of 85 subjects is the minimum requirement for a general validation, with 35 subjects per sub-group for specific population claims (e.g., pediatric, pregnancy, elderly, renal disease).
Executing a compliant clinical investigation under this standard requires meticulous logistical planning and strict adherence to protocol.