Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ISO 28057:2019 establishes standardized procedures for clinical dosimetry using solid thermoluminescence detectors for photon and electron radiations in radiotherapy. This international standard ensures consistency, accuracy, and traceability of TLD-based dose measurements across radiotherapy centers worldwide, supporting quality assurance programs, enabling intercomparison of treatment delivery accuracy, and providing the technical basis for remote dosimetry audit programs such as those operated by the IAEA and regional dosimetry networks.
The standard covers the fundamental principle of thermoluminescence dosimetry, where ionizing radiation excites electrons to metastable trapped states in crystalline lattice defects. During controlled heating, these trapped electrons return to ground state with emission of visible light proportional to the absorbed dose. ISO 28057 specifies requirements for commonly used TLD materials including LiF:Mg,Ti (TLD-100) and LiF:Mg,Cu,P (TLD-100H), detailing their annealing protocols, energy response characteristics, dose-response linearity range, fading rates, and optimal readout parameters. The standard also addresses the handling and storage conditions necessary to maintain TLD performance.
| TLD Material | Effective Atomic Number | Useful Dose Range | Fading Rate (annual) | Primary Application |
|---|---|---|---|---|
| LiF:Mg,Ti (TLD-100) | 8.2 (tissue-equivalent) | 10 μGy to 10 Gy | Approximately 5% per year | Clinical dosimetry, in-vivo patient verification |
| LiF:Mg,Cu,P (TLD-100H) | 8.2 (tissue-equivalent) | 1 μGy to 30 Gy | Less than 3% per year | Low-dose measurements, high-sensitivity audits |
| CaF2:Dy (TLD-200) | 16.3 (over-response to low energy) | 0.1 μGy to 10 Gy | Approximately 5% per year | Environmental radiation monitoring |
| Al2O3:C (OSL detector) | 10.2 (near-tissue equivalent) | 0.5 μGy to 20 Gy | Less than 1% per year | Personal dosimetry, reference audits |
ISO 28057 specifies detailed calibration procedures requiring direct traceability to primary standards dosimetry laboratories through accredited calibration chains. TLD readers must undergo regular quality control including light source stability tests, photomultiplier tube dark current monitoring, heating element temperature profile verification, and annealing oven uniformity checks. The complete measurement procedure encompasses TLD selection and screening, standardized annealing cycles, irradiation following clinical protocols, controlled readout with time-temperature profiles, signal processing with background subtraction, dose calculation using calibration curves with uncertainty propagation, and comprehensive uncertainty analysis following ISO/IEC Guide 98-3.
The standard supports critical clinical applications including in-vivo dosimetry for patient treatment verification, small-field dosimetry for stereotactic radiosurgery and IMRT/VMAT, output constancy checks for linear accelerators, brachytherapy source verification, and inter-institutional dosimetry audits. ISO 28057 serves as the reference standard for the IAEA/WHO TLD postal dose audit program, which has been instrumental in harmonizing radiotherapy dosimetry practices globally for over 50 years.
A comprehensive quality control program is essential for maintaining TLD dosimetry accuracy. ISO 28057 specifies daily, weekly, monthly, and annual QC procedures. Daily checks include reader warm-up stability test using a reference light source, background reading verification, and annealing oven temperature verification. Weekly procedures include reproducibility testing using a reference irradiation, and annual full recalibration at a standards laboratory. The standard provides action limits for each QC parameter and specifies corrective actions when limits are exceeded.
Inter-laboratory comparison programs are a key requirement of ISO 28057. Each laboratory must participate in at least one external dosimetry audit per year, typically through IAEA or regional audit networks. These comparisons verify that the laboratory’s TLD measurements are consistent with reference values and provide independent validation of the quality control program. Results are analyzed using standardized statistical methods including the well-established En number evaluation.
ISO 28057 specifies minimum training and competency requirements for personnel performing TLD dosimetry. Medical physicists and dosimetrists must demonstrate proficiency in TLD handling, readout procedures, calibration techniques, uncertainty analysis, and quality control procedures. The standard recommends documented initial training programs with competency assessment, continuing professional development with minimum annual hours, and periodic proficiency testing through blind audit irradiations.
TLD dosimetry following ISO 28057 plays a critical role in patient safety for radiotherapy. In-vivo dosimetry using TLDs provides independent verification that the prescribed radiation dose is delivered accurately, detecting errors in treatment planning, machine calibration, or patient setup that could otherwise lead to underdosing of tumors or overdosing of healthy tissues. Many radiotherapy centers have established routine in-vivo TLD programs for specific treatment sites such as breast, head and neck, and prostate treatments.