Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ISO 27893:2011 specifies the methodology for evaluating uncertainties of vacuum gauge calibrations performed by direct comparison with a reference gauge. This standard is fundamental to vacuum metrology, providing a consistent framework for calculating calibration uncertainty across the pressure range from 10E-7 Pa to atmospheric pressure. It applies to all vacuum gauge types including ionization gauges, capacitance manometers, Pirani gauges, and spinning rotor gauges.
Calibration by direct comparison involves simultaneously measuring the same pressure with both the gauge under test and a reference gauge with known traceability to primary pressure standards. The difference between readings, combined with all uncertainty components, constitutes the calibration result.
| Uncertainty Component | Type | Typical Magnitude | Evaluation Method |
|---|---|---|---|
| Reference gauge uncertainty | B | 0.5-5% of reading | Calibration certificate |
| Repeatability of test gauge | A | 0.2-2% of reading | Standard deviation |
| Resolution of test gauge | B | 0.1-1% FS | Digital resolution / 2 sqrt3 |
| Temperature effects | B | 0.1-0.5% per K | Temperature coefficient x dT |
| Gas composition effects | B | 1-10% for ion gauges | Ionization cross-section diff |
| Outgassing effects | B | 0.01-0.1% of reading | Rate-of-rise measurement |
The combined standard uncertainty uc is the root-sum-square of all components. Expanded uncertainty U = k x uc (k=2 for 95% confidence). Typical expanded uncertainties range from 1% for capacitance manometers at 10E2-10E5 Pa to 20% for hot cathode ionization gauges at 10E-5 Pa.
The standard specifies temperature control (+/- 1C), cleanliness (ISO Class 8 or better), and pumping system configuration. The reference gauge must have traceability to national standards with recalibration intervals not exceeding 24 months.