Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ISO 27891:2015 establishes standardized procedures for calibrating condensation particle counters (CPCs) used to measure aerosol particle number concentration. CPCs are fundamental instruments in aerosol science, capable of detecting nanoparticles down to 1 nm by condensing a working fluid onto particles to grow them to optically detectable sizes. Accurate calibration is essential for atmospheric aerosol research, air quality monitoring, and nanoparticle toxicology studies.
The standard specifies calibration using reference aerosol electrometry — the most direct primary standard for number concentration. The reference method uses a Faraday cage aerosol detector measuring electrical current from charged particles, providing SI-traceable measurement.
| Parameter | Method | Concentration Range | Uncertainty Target |
|---|---|---|---|
| Counting efficiency | Reference electrometer | 1E3-1E5 particles/cm3 | +/- 5% (k=2) |
| D50 detection size | Stepwise DMA scan | 1-100 nm | +/- 0.2 nm |
| Linear range | Dual CPC dilution | 1E0-1E7 particles/cm3 | +/- 10% upper limit |
| Coincidence correction | Pulse width analysis | >1E4 particles/cm3 | +/- 3% correction |
| Response time | Step concentration | T10-T90 | +/- 0.2 s |
The counting efficiency curve follows a sigmoidal function. D50 is defined as the particle diameter at 50% efficiency. For well-calibrated CPCs, efficiency plateaus at 95-100% for particles 2-3 times larger than D50. Calibration requires a minimum of five particle sizes spanning the D50 and plateau regions.
ISO 27891 protocols are referenced by global networks including GAW, ACTRIS, and IMPROVE. Network harmonization requires biennial recalibration with transfer standards ensuring inter-laboratory comparability within +/- 10%.