Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ISO 25902-1:2009 specifies the eddy-current examination method for detecting surface and near-surface discontinuities in seamless titanium pipes and tubes. As the first part of the ISO 25902 series on non-destructive testing of titanium products, this standard addresses the critical need for reliable quality control in the production of titanium tubing used in aerospace, chemical processing, medical implants, and power generation applications.
Titanium presents unique challenges for eddy-current testing. Its relatively low electrical conductivity (approximately 1.8% IACS, compared to copper’s 100% IACS) and non-ferromagnetic nature mean that standard eddy-current techniques must be carefully optimized. The standard provides specific guidance on test frequencies, coil configurations, and sensitivity settings tailored to titanium’s material properties.
| Parameter | Specification | Notes |
|---|---|---|
| Applicable diameter range | Per agreement between parties | Typically 6 mm to 100+ mm OD |
| Minimum wall thickness | Not specified (detection limited) | Depends on frequency and coil |
| Material conductivity | ~1.8% IACS | Requires low frequency optimization |
| Reference notch depth | Per agreement (typical 5-10% wall) | Longitudinal and transverse |
| Test frequency range | Typically 10 kHz – 2 MHz | Optimized for titanium properties |
| Coil type | Encircling (feed-through) or probe | Depends on tube dimensions |
The apparatus requirements (Clause 5) cover the eddy-current test instrument, coils, and associated mechanical handling equipment. The instrument must provide adjustable frequency, gain, phase analysis, and alarm threshold settings. For titanium tubes, the relatively low conductivity means that test frequencies are typically lower than those used for copper or aluminium tubing to achieve adequate penetration depth.
The standard defines three fundamental coil configurations:
Encircling coils (feed-through) — The tube passes through a coil that generates a circumferential magnetic field. This configuration detects longitudinal discontinuities and provides the fastest inspection speed for production-line testing.
Probe coils (internal or external) — A small probe is scanned over the tube surface. This provides higher sensitivity for small defects and can detect discontinuities in any orientation, but at slower inspection speeds.
Sector coils — Covering a partial circumference, these are used for specific applications where access is limited or targeted detection is required.
The reference piece (Clause 6) is a critical element of the examination system. It consists of a tube sample of the same material, dimensions, and surface condition as the production tubes, containing artificial defects for calibration. The standard specifies that reference standards shall contain notches with dimensions agreed upon between the purchaser and supplier. Typical notch depths range from 5% to 20% of the wall thickness, with both longitudinal and transverse orientations to ensure detection of defects in any direction.
| Reference Defect Type | Typical Dimensions | Purpose |
|---|---|---|
| Longitudinal notch (outer surface) | 5-20% wall depth, length ≥ 2× width | Detection of longitudinal flaws |
| Transverse notch (outer surface) | 5-20% wall depth | Detection of transverse/circumferential flaws |
| Longitudinal notch (inner surface) | 5-20% wall depth | Detection of internal surface flaws |
| Through-hole (drilled) | 0.5 – 2.0 mm diameter | General sensitivity verification |
The examination procedure (Clause 7) details the step-by-step methodology for conducting eddy-current tests on titanium tubes. Standard reference frequency selection is based on the material’s electromagnetic properties and the tube dimensions. The standard depth of penetration (δ) for eddy currents in titanium follows the classical skin effect equation:
δ = 1 / √(π f μ σ)
Where f is frequency, μ is magnetic permeability, and σ is electrical conductivity. For titanium (non-ferromagnetic, μ = μ₀), this simplifies to show that penetration depth is inversely proportional to the square root of frequency. Typical test frequencies for titanium tube examination range from 10 kHz to 500 kHz for encircling coils, with lower frequencies providing deeper penetration but reduced sensitivity to small surface defects.
The standard specifies that the confirmation of sensitivity (7.4) must be performed before each inspection run, at regular intervals during production, and after any significant change in test conditions. Sensitivity is verified by passing the reference standard through the test system and confirming that all artificial defects produce signals above the alarm threshold. If sensitivity drifts below the required level, the examination system must be recalibrated and all tubes tested since the last valid calibration must be re-examined.
Acceptance criteria (Clause 8) are determined by agreement between the purchaser and supplier, based on the tube’s intended application. For critical applications such as aerospace hydraulic lines, very stringent criteria apply — typically rejecting any defect indication exceeding 5% of wall thickness. For less critical applications such as general industrial heat exchanger tubing, more lenient criteria of 10-20% wall thickness may be acceptable.
The standard distinguishes between:
Conditional acceptance — Tubes with indications exceeding the alarm threshold may be accepted if the defect is removed by grinding or other means, provided that the remaining wall thickness remains within specified tolerances.
Rejection — Tubes with defects that cannot be removed or that reduce wall thickness below the minimum specified value are rejected.
The test report (Clause 9) must include: tube identification, material specification, dimensions, test instrument details, frequency, coil type, reference standard calibration data, test results summary, acceptance criteria applied, operator identification, and date of examination. This documentation provides traceability essential for quality management systems and regulatory compliance.
For engineers designing titanium tube inspection systems, several practical considerations emerge from the standard. First, surface condition significantly affects eddy-current response — rough surfaces, surface oxidation (common in titanium), and residual lubricants can produce noise signals that mask genuine defects. Proper surface preparation and signal filtering are essential.
Second, end effects — the disruption of eddy-current fields at tube ends — limit the inspectable length. Typically, the first and last 20-50 mm of each tube length cannot be reliably examined with encircling coils. For critical applications, supplementary probe inspection of end regions may be necessary.
Third, lift-off effects — variations in coil-to-tube spacing — are a major source of noise in eddy-current testing. For titanium tubes, where signal amplitudes from genuine defects may be relatively small, maintaining consistent probe position is particularly important. Mechanized scanning systems with precision guidance provide significantly better signal-to-noise ratios than manual inspection.