ISO 17827-2:2017 — Determination of Particle Size Distribution for Uncompressed Solid Biofuels (Vibrating Screen Method for Sieve Apertures ≤3.15 mm)

Technical insights into the international standard for fine particle sizing in biomass fuels, adopted as CAN/CSA-ISO 17827-2:17

Scope and Introduction

ISO 17827-2:2017, adopted in Canada as CAN/CSA-ISO 17827-2:17, specifies a method for determining the particle size distribution (PSD) of uncompressed solid biofuels using a vibrating screen with sieve apertures of 3.15 mm and below. This standard is part of the ISO 17827 series, which addresses PSD for uncompressed fuels under different size ranges. Part 2 focuses on finer particles (≤3.15 mm), complementing Part 1 which covers larger fractions (≥3.15 mm). The method applies to materials such as crushed wood pellets, sawdust, hog fuel, olive residues, and other particulate biomass that have a significant proportion of particles smaller than 3.15 mm.

The vibrating screen method described in ISO 17827-2 employs a set of sieves placed in a stack, subjected to controlled vibration to separate particles by size. The standard is essential for quality control in biofuel production, combustion efficiency optimization, and compliance with fuel classification systems (e.g., ISO 17225 series). By harmonizing test procedures internationally, ISO 17827-2 enables reliable comparison of PSD results across laboratories and supply chains, directly supporting trade and utilization of solid biofuels.

Why it matters: Consistent particle size distribution is critical for fuel feeding systems, combustion behavior, and emission control. ISO 17827-2 provides a reproducible, standardized method to assess fine particle fractions, which significantly influence fuel handling and conversion performance.

Technical Requirements

Apparatus and Sieves

The standard mandates the use of test sieves conforming to ISO 3310-1 (woven wire cloth) or ISO 3310-2 (perforated metal plate). Sieve aperture sizes shall be selected to cover the expected particle size range, typically including 3.15 mm, 2.00 mm, 1.00 mm, 0.500 mm, 0.250 mm, and a bottom pan. Each sieve must have a nominal diameter of 200 mm (or 300 mm for larger samples) and comply with the tolerances specified in ISO 3310. The recommended aperture increments follow a ratio of √2 from the 3.15 mm reference. Table 1 lists suggested sieve apertures and corresponding tolerances.

Table 1 — Typical sieve apertures and tolerances per ISO 17827-2:2017
Nominal aperture (mm) Permissible deviation (ISO 3310‑1) (μm) Intermediate tolerance (μm) Application range
3.15 ±140 +190 Coarse end of fine fraction
2.00 ±100 +140 Medium fraction
1.00 ±60 +80 Medium fraction
0.500 ±35 +45 Fine fraction
0.250 ±18 +25 Very fine fraction
Pan (0) Collector for material passing finest sieve

A vibrating sieve shaker capable of delivering a vertical vibration amplitude of 0.5 mm to 3 mm (at 50 Hz) or equivalent shall be used. The vibration parameters must be adjustable to achieve optimal sieving efficiency without damaging the sieves or altering the sample. Additional equipment includes a balance with a precision of 0.1 g, drying oven (set at 105 °C for moisture determination), and sample dividers (e.g., riffle boxes) to obtain representative subsamples.

Sample Preparation

The test sample must be taken from a larger lot following the principles of ISO 18135 (Sampling of solid biofuels) and ISO 14780 (Sample preparation). The sample shall be air-dried to a moisture content of ≤15 % by mass (wet basis) to avoid clogging of sieves and to ensure free-flowing behavior. If the as-received moisture exceeds this threshold, the sample must be dried in an oven at 50 °C until the desired moisture level is reached, taking care not to devolatilize the fuel. The mass of the test portion (Mr) depends on the estimated nominal top size of the material (as defined in ISO 17827-1):

  • For materials where ≥90 % passes the 3.15 mm sieve, a minimum sample mass of 100 g is required.
  • If the material is coarser (retained fraction on 3.15 mm >10 %), the test portion must be at least 200 g to ensure representativeness.

The sample shall be divided into two replicates for duplicate determinations; the mean of both results is reported as the final PSD.

Procedure

  1. Assemble the sieve stack in ascending order of aperture (largest at top, pan at bottom).
  2. Weigh the empty sieves and pan. Place the test portion on the top sieve.
  3. Install the lid and secure the stack on the vibrating shaker.
  4. Set the vibration amplitude and frequency to achieve a vertical motion of 1.5 mm–2.5 mm at 50 Hz (typical). The duration of sieving is initially 10 min for most materials; longer times (up to 20 min) may be required for high‑fines content.
  5. After sieving, carefully disassemble the stack. Weigh each sieve and the pan with the retained material. Calculate the mass retained on each sieve by subtracting the initial tape weight.
  6. Optionally perform a hand-sieving check (as per Annex A of the standard) to verify the end point: further sieving for 2 min should result in less than 0.5 % by mass passing the finest sieve.
  7. Repeat with the second replicate.
TIP: To avoid static electricity effects and improve sieving efficiency for very fine particles (<0.25 mm), consider adding a drop of an antistatic agent (e.g., 0.1 % solution of surfactant) to the sample before sieving, as recommended in ISO 17827-2:2017 Annex B for problematic materials.

Calculation and Reporting

For each sieve size i, compute the mass fraction Ri (mass retained on sieve i) as a percentage of the total test mass:

Ri = (mi / Mr) × 100

where mi is the mass retained on sieve i (in grams) and Mr is the total mass of the test portion (in grams). The cumulative undersize (Pi) is then derived: Pi = 100 − ΣiRi. Results are reported as the average of the two replicate determinations, rounded to the nearest 0.1 % for each fraction. The report must include: the moisture content of the sample, the sieve apertures used, the sieving time and amplitude, and any deviations from the standard procedure.

WARNING: Moisture content above 15 % can cause agglomeration of particles and block sieve apertures, leading to biased results. Always dry the sample to the specified level before testing, and confirm the final moisture using a calibrated moisture analyzer.

Implementation Highlights

ISO 17827-2:2017 provides a critical tool for biofuel producers and users to characterize the fine fraction of their fuels. The standard is often referenced in fuel quality specifications (e.g., for ISO 17225‑2 for wood pellets, and ISO 17225‑4 for wood chips) to ensure that the amount of fines (material passing a 1 mm or 0.5 mm sieve) is within acceptable limits for the intended combustion technology.

Key differences from ISO 17827‑1 (which covers coarse particles ≥3.15 mm) include:

  • The use of a vibrating screen rather than a horizontal or horizontal‑rotational sieve motion, because fine particles are more effectively separated by vertical vibration.
  • Sieve sizes are limited to ≤3.15 mm, but the method may be extended to include a 4 mm or 5.6 mm sieve at the top if a small fraction of oversize is present (per the scope extension in Annex C).
  • Greater emphasis on avoiding electrostatic charging and moisture effects.

The method is suitable for materials with a nominal top size (after cutting or crushing) of up to 3.15 mm, such as hog fuel, sawdust, and certain agricultural residues. In cases where the sample contains an appreciable amount of wood bark or fibrous material, a 1‑minute hand‑sieving step after the mechanical sieving is permitted to free trapped particles. The flexibility of the standard allows laboratories to adjust sieving parameters (within defined limits) to suit the specific fuel type, as long as the chosen conditions are documented and validated.

COMMON PITFALL: Overloading the top sieve (exceeding the recommended sample mass) reduces sieving efficiency and can damage the sieve cloth. Always respect the sample mass guidelines in Clause 7.3 of the standard. Mass retained on any sieve after sieving should not exceed the values given in ISO 3310‑1 for that sieve diameter; if it does, split the sample and run multiple sieving passes.

Compliance Notes

ISO 17827-2:2017 is a Tier 1 test method for the determination of particle size distribution of solid biofuels under the ISO 17225 series. Compliance with the standard is required for certification to the quality classes defined in ISO 17225 (e.g., A1, A2, B for pellets). Laboratories that carry out PSD testing should operate under a quality management system such as ISO 17025, and use verified reference sieves that are calibrated at regular intervals.

In Canada, the standard has been adopted as CAN/CSA‑ISO 17827‑2:17 by the Canadian Standards Association. This adoption is identical to the international version, with no deviations. Compliance with CAN/CSA‑ISO 17827‑2:17 may be required by regulatory bodies or by contractual agreements for biomass supply to Canadian power plants and district heating systems. Similarly, other countries may adopt the standard regionally, often with the same technical content.

When reporting test results for compliance purposes, the laboratory must include the complete standard reference (e.g., “CAN/CSA‑ISO 17827‑2:17”) and state the moisture content of the sample at the time of testing. Any method modifications (e.g., extended sieving time, use of wet‑sieving for sticky materials) must be clearly noted and justified. Inter‑laboratory comparisons are recommended to ensure reproducibility, especially when the PSD is used to verify fuel quality guarantees.

Compliance benefit: Uniform particle size data generated according to ISO 17827‑2 enables transparent communication between fuel producers, traders, and end users. It reduces disputes related to fines content and helps maintain consistent boiler operation, lower emissions, and reduced maintenance costs.

Frequently Asked Questions

Q: What is the difference between ISO 17827‑1 and ISO 17827‑2?
A: ISO 17827‑1 covers the method for determining particle size distribution using sieves with apertures of 3.15 mm and above (coarse fraction), typically using a horizontal‑rotational sieve shaker. ISO 17827‑2 covers apertures ≤3.15 mm (fine fraction) and prescribes a vertical vibration (vibrating screen) method. They are complementary; together they cover the full size range for uncompressed solid biofuels. If a fuel sample spans both ranges, the two standards are applied sequentially to overlapping sieve sets.
Q: Can I use ISO 17827‑2 for materials other than solid biofuels?
A: The standard is specifically developed for solid biofuels (e.g., wood pellets, chips, briquettes, and agricultural residues). However, the method can be applied to other granular materials (e.g., biomass pellets, wood flour) if the analyst validates that the procedure yields repeatable and accurate results. The standard states that it may be used for other particulate materials as long as the applicability is verified.
Q: What is the recommended sieving time?
A: The initial sieving time shall be 10 minutes. If after 10 minutes the mass passing the finest sieve in a 2‑minute hand‑sieving check exceeds 0.5 % of the total sample mass, the mechanical sieving is continued in increments of 2 minutes until the end point is reached. The maximum recommended sieving time is 20 minutes, after which the sieves should be inspected for blinding or blockage.
Q: How should I store and clean the sieves after use?
A: Clean sieves carefully with a soft brush to remove trapped particles. Never use force or sharp objects that may distort the wire mesh. Store sieves in a dry, dust‑free environment. Periodically inspect the sieve cloth for damage and recalibrate according to the laboratory’s standard operating procedures and the sieve manufacturer’s recommendations.

Last updated 2026 — This article provides general technical guidance; always refer to the official ISO or CSA publication for complete requirements.

📥 Standard Documents Download

🔒
Please wait 10 seconds, the download links will appear after the ad loads

Leave a Reply

Your email address will not be published. Required fields are marked *