Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ISO 15197‑15 (CAN/CSA‑ISO 15197:15) is the Canadian adoption of the international standard ISO 15197:2013, “In vitro diagnostic test systems — Requirements for blood‑glucose monitoring systems for self‑testing in managing diabetes mellitus.” It applies to blood‑glucose monitoring systems (BGMS) intended for use by laypersons at home, covering the entire system: meter, test strips, control solutions, and all accessories.
The standard defines essential performance requirements to ensure that self‑monitored glucose results are sufficiently accurate for clinical decision‑making. It also outlines evaluation procedures for manufacturers and conformity‑assessment bodies. Health Canada recognizes ISO 15197‑15 as the benchmark for pre‑market review of BGMS sold in Canada, aligning with the international consensus on quality and safety.
The core of ISO 15197‑15 is the accuracy requirement expressed as the agreement between BGMS results and a reference method (e.g., hexokinase or glucose oxidase on a clinical analyzer). The acceptance criteria differ by glucose concentration range:
| Glucose Concentration | Accuracy Criterion (95 % of results shall fall within) |
|---|---|
| < 5.6 mmol/L (100 mg/dL) | ±0.83 mmol/L (15 mg/dL) of the reference value |
| ≥ 5.6 mmol/L (100 mg/dL) | ±15 % of the reference value |
Additionally, in the lower concentration range (< 5.6 mmol/L), at least 99 % of results must fall within ±1.67 mmol/L (30 mg/dL) or ±30 % (whichever is greater). The standard mandates that accuracy be verified by at least 100 capillary blood samples from a diverse subject population, tested in triplicate with at least three different reagent lot numbers.
ISO 15197‑15 requires manufacturers to evaluate the effect of endogenous and exogenous interfering substances. The user‑performed test must maintain accuracy in the presence of:
An essential requirement is demonstration that lay users can obtain accurate results under real‑life conditions. The standard calls for a clinical study involving at least 100 untrained subjects who read the instructions and perform a capillary blood test on themselves. The acceptance criterion is the same as for system accuracy (see table above). If the system includes alternate‑site testing (e.g., forearm), similar studies must be performed for each site.
To comply with ISO 15197‑15, manufacturers must design BGMS that minimize systematic biases and random errors. The development process typically includes:
In Canada, Health Canada recognizes ISO 15197‑15 as a special control for class III BGMS. Manufacturers submit a pre‑market application (Medical Device License) that includes the full evaluation dossier: accuracy studies, interference tests, user study, and stability data. A conformity‑assessment body (e.g., CSA Group) may audit the manufacturer’s quality system and review the technical documentation against the standard’s requirements.
Manufacturers should also note that ISO 15197:2013 (and thus ISO 15197‑15) is under periodic review. A revision is expected in the coming years, potentially tightening the accuracy criteria (e.g., moving to ±10 % above 5.6 mmol/L). Staying informed of the standard’s evolution is essential for long‑term product compliance.
— Updated 2026 —