Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
IEC TS 62903:2018 specifies a precise measurement method for determining the electroacoustical parameters and acoustic output power of spherically curved ultrasonic transducers using the self-reciprocity technique. This technical specification addresses a critical need in therapeutic ultrasound applications, including high-intensity focused ultrasound (HIFU) for tumour ablation, ultrasound physiotherapy, and lithotripsy, where accurate knowledge of acoustic output power is essential for treatment efficacy and patient safety.
The self-reciprocity technique leverages the fundamental principle of acoustical reciprocity for linear, passive, and reversible transducers. For a spherically focused transducer, the standard defines a measurement configuration where the transducer is driven electrically to generate an acoustic field, which is reflected by a planar reflector positioned at the focal plane and received by the same transducer operating in reception mode. The electrical transfer impedance measured at the transducer terminals encodes both the transmit and receive characteristics of the device.
By combining measurements of the electrical impedance, the echo response from the reflector, and the theoretical diffraction correction for a spherical wave, the transducer’s open-circuit sensitivity and acoustic output power can be calculated without external calibration. The standard specifically addresses the case where the transducer radius a and focal length F satisfy the relation F ≥ 2a²/λ (where λ is the acoustic wavelength), ensuring valid far-field conditions at the focal plane.
| Measurement Parameter | Symbol | Required Accuracy | Measurement Principle |
|---|---|---|---|
| Electrical input impedance | Ze | ±2% | Network analyser or impedance bridge |
| Echo signal voltage | Vecho | ±3% | Digitising oscilloscope (8-bit min) |
| Reflector distance | d | ±0.1 mm | Laser interferometry or mechanical gauge |
| Water temperature | T | ±0.5 °C | Calibrated thermocouple (type K or T) |
| Acoustic output power | Pac | ±10% (k=2) | Self-reciprocity calculation |
The standard defines the measurement system configuration in detail. A water tank with dimensions at least 5× the transducer diameter in each direction is required to avoid boundary reflections. Acoustic absorbing material must line all tank walls except the transducer mounting surface. The reflector shall be a polished stainless steel or glass plate with surface flatness better than λ/10 at the operating frequency and positioned perpendicular to the acoustic axis within 0.5°.
The electrical measurement system must include a function generator capable of producing tone bursts with controlled amplitude, duration, and repetition rate. The standard recommends burst durations of 10–50 cycles with a duty cycle below 1% to avoid thermal effects in the transducer. A diplexer or directional coupler separates the transmitted and received signals, with isolation exceeding 40 dB between transmit and receive paths.
The acoustic output power Pac is derived from the measurement data using the self-reciprocity equation, which combines the transmitted voltage, electrical impedance, focal length, water properties, and diffraction correction. The standard provides a comprehensive uncertainty budget analysis, estimating expanded uncertainty (k=2) of ±10% for acoustic output power measurement, with the dominant contributions coming from echo signal amplitude measurement and transducer positioning alignment.
The standard also specifies validation procedures using a radiation force balance (RFB) method as an independent cross-check. The RFB method measures acoustic power by detecting the radiation force exerted on a target, providing a complementary measurement with different systematic error sources. Agreement between the two methods within the combined uncertainty bounds serves as validation of the measurement setup.