IEC TR 62331 is a Technical Report that describes the principles, instrumentation, and measurement techniques of pulsed field magnetometry (PFM) for characterizing permanent magnetic materials. Unlike conventional hysteresisgraph or permeameter methods that require slowly varying fields, PFM generates a complete hysteresis loop in milliseconds by applying a brief but intense magnetic field pulse to the test specimen.
The fundamental principle involves discharging a capacitor bank through a magnetizing solenoid to generate a pulsed magnetic field of sufficient amplitude to saturate the test material. During the pulse rise and decay, the magnetic polarization J (or magnetization M) and the magnetic field strength H are measured using specially designed pick-up coils and transient digitizers. The resulting J(H) and B(H) hysteresis loops provide all essential magnetic parameters including remanence (Br), coercivity (HcB, HcJ), and maximum energy product (BHmax).
The field generator subsystem consists of three critical components: a high-voltage power supply, an energy storage capacitor bank, and a magnetizing solenoid. The power supply charges the capacitor bank to voltages ranging from hundreds to several thousand volts, depending on the required field amplitude. For measuring high-coercivity rare-earth magnets, peak fields of 3000-5000 kA/m (approximately 4-6 T in free space) are typically required.
| Field Generation Method | Pulse Shape | Peak Field | Pulse Duration | Application |
|---|---|---|---|---|
| Capacitor discharge (sine wave) | Decaying sine | ≤ 5000 kA/m | 2-20 ms | General purpose, most common |
| Unidirectional half-sine | Half-sine pulse | ≤ 3000 kA/m | 5-50 ms | Low coercivity materials |
| Unidirectional decaying | RC discharge | ≤ 7000 kA/m | 1-10 ms | Ultra-high coercivity materials |
Accurate measurement of J and H requires carefully designed pick-up coils. The polarization sensor (J coil) is wound closely around the test specimen to capture the magnetic flux from the sample. The magnetic field strength sensor (H coil) is positioned adjacent to the sample to measure the applied field. The standard discusses various coil configurations, including concentric coils, tangential pairs, and printed-circuit board coils, each with specific advantages for different specimen geometries.
The induced voltages from pick-up coils are proportional to dJ/dt and dH/dt, requiring integration to obtain J and H values. The standard describes both analogue integration (using operational amplifier integrators) and numerical integration (after high-speed digitization). For numerical integration, the digitization rate must be sufficiently high (typically ≥ 1 Msample/s per channel) to capture the rapid signal changes during the pulse.
| Processing Step | Purpose | Critical Parameters |
|---|---|---|
| Drift correction | Remove integrator drift and offset | Pre-pulse baseline duration |
| Eddy current correction | Compensate for conductive specimen eddy currents | Conductivity, specimen dimensions |
| Demagnetization correction | Account for specimen shape effects | Demagnetization factor (Nd) |
| Temperature correction | Adjust for pulse heating of specimen | Specific heat, temperature coefficient |
| Calibration scaling | Convert digitizer units to SI magnetic units | Calibration constants |
The standard acknowledges that magnetic viscosity (time-dependent magnetization changes) can cause significant differences between PFM and DC measurements, particularly for materials with high intrinsic coercivity. The standard provides guidance on comparing PFM results with conventional measurements and interpreting the differences in terms of thermally activated magnetization reversal processes.
The Technical Report presents extensive comparative measurement data between PFM systems and conventional methods. For large permanent magnet specimens, PFM measurements show excellent agreement with permeameter results within ± 2% for Br and ± 3% for HcJ. For small specimens, comparison with extraction method magnetometers (EMM) shows similar agreement, validating PFM as a reliable measurement technique.
| Parameter | PFM vs. Permeameter (Large) | PFM vs. EMM (Small) |
|---|---|---|
| Remanence (Br) | ± 2% | ± 2% |
| Coercivity (HcB) | ± 2% | ± 2.5% |
| Intrinsic coercivity (HcJ) | ± 3% | ± 3% |
| Maximum energy product (BHmax) | ± 4% | ± 4% |
| Measurement time | < 1 s | < 1 s |
At the time of publication (2005), pulsed field magnetometry was still evolving as a measurement technique, and consensus on all details of standardized test procedures had not yet been reached. The Technical Report format allows practitioners to benefit from documented best practices while the methodology continues to develop.
The main uncertainty contributions include: eddy current correction accuracy (±1%), calibration uncertainty (±0.5%), temperature effects (±0.5%), signal-to-noise ratio (±0.3%), and demagnetization correction (±0.3%). The combined expanded uncertainty (k=2) is typically ±3-5% depending on material and specimen geometry.
Yes, but with limitations. Soft magnetic materials (low coercivity) require very short pulse durations and careful control of eddy currents. The standard notes that PFM is best suited for hard (permanent) magnet materials. For soft materials, conventional DC or AC methods (e.g., IEC 60404 series) are generally preferred.
The pulse shape directly affects eddy current magnitude (faster pulses induce stronger eddy currents), signal-to-noise ratio (slower pulses give better SNR but longer measurement time), and adiabatic heating (shorter pulses reduce heating). The optimal pulse shape depends on the material’s electrical conductivity, coercivity, and thermal properties.