IEC 62311 is the horizontal standard for evaluating electronic and electrical equipment with respect to human exposure to electromagnetic fields. It serves as the product-family standard when no dedicated product-specific EMF standard exists. The standard applies to equipment intended for both residential and occupational environments, including consumer electronics, telecommunications equipment, industrial machinery, medical devices, and household appliances. The assessment covers all relevant exposure metrics: induced electric field and current density for low frequencies (below 10 MHz), specific absorption rate (SAR) for intermediate frequencies (100 kHz to 10 GHz), and incident power density for high frequencies (above 10 GHz).
The standard aligns with the exposure limits established by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) and the IEEE C95.1 standard, providing a practical compliance pathway that uses either computational simulation (e.g., finite-difference time-domain/ FDTD methods) or physical measurements (e.g., isotropic field probes, SAR measurement systems) depending on the equipment type, frequency range, and available resources.
IEC 62311 introduces a three-tier assessment approach that progressively increases in complexity:
When performing physical measurements, the standard specifies probe positioning, environmental conditions, and uncertainty analysis per ISO/IEC Guide 98-3 (GUM). For computational methods, it requires validated solver codes, appropriate mesh resolution (at least 10 cells per wavelength), and convergence criteria. The standard also addresses the increasingly important issue of multiple sources and cumulative exposure—products with multiple transmitters require combined assessment.
| Frequency Range | Exposure Metric | Basic Restriction (General Public) | Typical Assessment Method |
|---|---|---|---|
| 0 Hz (static) | Flux density B | 400 mT (static magnetic) | Gaussmeter measurement |
| 1 Hz – 100 kHz | Induced E-field (Ei) | 1.1 – 4.0 V/m (50/60 Hz) | FEM simulation or field measurement |
| 100 kHz – 10 MHz | SAR (transition zone) | 0.08 W/kg (whole-body) | Hybrid SAR + E-field measurement |
| 10 MHz – 10 GHz | SAR | 2.0 W/kg (10g avg, localized) | Dosimetric assessment per IEC 62209 |
| 10 GHz – 300 GHz | Power density (Seq) | 10 W/m² | Isotropic probe with E/H field sensors |
Measurement uncertainty is a central theme in IEC 62311. The standard requires that the expanded measurement uncertainty (k = 2, 95% confidence level) be below 30% for SAR assessments and below 4 dB for power density measurements. When the measured value plus the expanded uncertainty exceeds the limit, the assessment is considered non-compliant. This conservative approach ensures that compliance decisions account for the inherent variability in both measurement instrumentation and human anatomical models.
The standard also addresses product categories that are often overlooked—wireless power transfer systems (Qi charging, resonant chargers) operating at tens to hundreds of kHz require assessment of induced currents and contact currents, which are not well covered by higher-frequency SAR standards. For these products, IEC 62311 specifies the use of anatomically realistic numerical models and induced E-field averaging over 2 mm × 2 mm × 2 mm voxels in the central nervous system tissues.
Basic restrictions are the actual biophysical limits on exposure (e.g., SAR, induced E-field) that are derived from established health effects. Reference levels are more easily measured field strength values (E-field, H-field, power density) that are set well below the basic restrictions to provide a safety margin. If the equipment’s emissions are below the reference levels, the basic restrictions are automatically met.
Yes, but for wireless transmitters, product-specific standards such as IEC 62209 (SAR for body-mounted devices) or IEC 62232 (RF field strength) take precedence. IEC 62311 serves as the fallback standard when no product-specific standard exists, or as a framework for assessing products with multiple transmitter types simultaneously.
Re-assessment is required whenever the product design changes in a way that affects EMF characteristics—changes to the antenna structure, output power, operating frequency, enclosure materials, or typical use configurations all trigger re-assessment. There is no periodic re-assessment requirement for unchanged products.
The standard requires expanded uncertainty (k = 2) of ≤ 30% for SAR measurements. When using computational methods, similar or better uncertainty must be demonstrated through validation studies comparing simulation results against reference measurements in standard phantoms.