Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Medical ultrasound imaging relies on pulse-echo scanners that transmit acoustic pulses into tissue and reconstruct images from reflected echoes. The diagnostic utility of an ultrasound system depends on several key performance parameters, including maximum depth of penetration, spatial resolution, contrast resolution, and dynamic range. Without standardized measurement methods, comparing systems from different manufacturers or evaluating performance degradation over time becomes subjective and unreliable.
IEC 61391-2-2010 addresses this need by defining reproducible test methods using tissue-mimicking phantoms. The standard is part of the IEC 61391 series, which covers ultrasonic pulse-echo scanners. Part 2 specifically focuses on two fundamental measurements: maximum depth of penetration (the greatest depth from which detectable echoes can be obtained) and local dynamic range (the ratio of the largest to smallest detectable echo signal at a given depth).
The maximum depth of penetration is a critical metric that determines how deeply a scanner can visualize structures within the body. This parameter depends on the transmit power, transducer frequency, receiver gain, and the attenuation characteristics of tissue. Lower-frequency transducers (2-5 MHz) typically achieve greater penetration depths (up to 30 cm in abdominal imaging), while higher-frequency transducers (7-15 MHz) offer better resolution but shallower penetration (typically 4-10 cm for vascular or small-parts imaging).
The standard specifies a test procedure using a tissue-mimicking phantom with known attenuation coefficient (typically 0.5 dB/cm/MHz, matching the average attenuation of soft tissue). The test involves acquiring an image of the phantom at maximum system gain and measuring the depth at which the signal-to-noise ratio falls below a defined threshold. The phantom contains a series of wire targets or cyst-like structures at known depths to provide reference echo signals.
| Parameter | Specification per IEC 61391-2 |
|---|---|
| Phantom Attenuation | 0.5 ± 0.05 dB/cm/MHz at 22 °C |
| Phantom Speed of Sound | 1540 ± 10 m/s at 22 °C |
| Measurement Depth Range | 0 to 30 cm (depending on transducer) |
| Threshold Definition | Echo amplitude > noise floor + 6 dB |
| Transducer Frequency Range | 1 to 15 MHz (covered by the standard) |
| Test Environment Temperature | 22 ± 3 °C |
Local dynamic range characterizes the scanner’s ability to display simultaneous signals of vastly different amplitudes — for example, a strong echo from a tissue boundary next to a weak echo from diffuse scattering within the same region. A wider dynamic range produces images with smoother texture and better differentiation of subtle tissue variations, but may reduce perceived contrast. The standard specifies measuring the dynamic range at a specific depth using a phantom with calibrated echo targets of varying reflectivity.
The measurement procedure involves scanning a phantom that contains targets with known echo amplitude differences (typically in 5 dB or 10 dB steps). The scanner’s display is evaluated to determine the range of echo amplitudes that can be simultaneously visualized without changing system settings. The local dynamic range is expressed in decibels (dB), with typical values for modern ultrasound systems ranging from 50 to 70 dB. Systems intended for cardiac imaging typically require higher dynamic range to visualize blood-tissue boundaries, while vascular systems may prioritize resolution over dynamic range.
The accuracy of both penetration depth and dynamic range measurements depends critically on the quality of the test phantom. IEC 61391-2 specifies detailed requirements for phantom construction, including the acoustic properties of the tissue-mimicking material (TMM). The phantom must have a uniform speed of sound of 1540 m/s (the average speed in soft tissue), an attenuation coefficient of 0.5 dB/cm/MHz, and a backscatter coefficient representative of human liver parenchyma.
| Phantom Property | Requirement | Tolerance |
|---|---|---|
| Speed of Sound | 1540 m/s | ± 10 m/s |
| Attenuation Coefficient | 0.5 dB/cm/MHz | ± 0.05 dB/cm/MHz |
| Backscatter Coefficient | Liver-equivalent | ± 3 dB |
| Non-linearity Parameter (B/A) | 7.0 ± 1.0 | Per manufacturer specification |
| Operating Temperature Range | 10 to 35 °C | ± 1 °C stability |
| Phantom Shelf Life | ≥ 2 years (sealed) | Per manufacturer |
When implementing test procedures per IEC 61391-2, engineers should consider several practical factors. First, phantom temperature directly affects the speed of sound and attenuation coefficient. The phantom must be allowed to stabilize at the test temperature (typically 22 °C) for at least 24 hours before measurement, as thermal gradients within the phantom material can produce measurement errors of 5% or more.
Second, transducer coupling to the phantom surface must be consistent. Use a standardized coupling gel and apply consistent pressure. Variations in coupling can change the apparent penetration depth by 1-2 cm. Many test laboratories use a fixture with a weighted transducer holder to ensure reproducible coupling pressure.
Third, the system settings used for measurement must be recorded and maintained for all periodic tests. Key parameters include transmit power, gain (overall and time-gain compensation), dynamic range setting, and post-processing curve. Any change in these settings invalidates the baseline comparison. The standard recommends establishing a “reference system configuration” that is used for all acceptance and consistency tests.
IEC 61391-1 covers general requirements for measuring the performance of pulse-echo ultrasound scanners, including test setup, phantom specifications, and general measurement principles. IEC 61391-2 specifically details the methods for measuring maximum depth of penetration and local dynamic range, providing step-by-step procedures for these two critical parameters.
The standard recommends performing the full set of measurements at acceptance (when a new system is installed), annually thereafter, and after any major repair or component replacement. Monthly quick checks of penetration depth using a simplified protocol are recommended for clinical quality assurance programs.
Yes, the principles apply to 3D and 4D systems, but additional considerations are needed for volumetric measurements. The 2D measurement methods form the foundation, and extended procedures for volumetric systems are covered in other parts of the IEC 61391 series and related standards such as IEC 61390-3.
For abdominal imaging with a 3.5 MHz convex transducer, typical penetration depth is 20-30 cm. For cardiac imaging with a 2.5 MHz phased array, penetration is 15-25 cm. For vascular imaging with a 7.5 MHz linear array, penetration is 4-8 cm. These values depend on system design, transducer technology, and phantom conditions.