Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
IEC 61223, developed by IEC Subcommittee 62B (Diagnostic imaging equipment), is a comprehensive series of standards providing systematic quality assurance testing frameworks for various imaging modalities. The series defines three test types: acceptance tests conducted after equipment installation or major repair to verify contractual specifications; status tests performed periodically for comprehensive performance assessment and trend detection; and routine tests (constancy tests) performed daily or weekly as quick checks to confirm stable equipment operation.
The series is structured into multiple parts covering different modalities: IEC 61223-1 covers general requirements and terminology; IEC 61223-2 series covers X-ray radiography and fluoroscopy equipment (parts -2-1 through -2-11); IEC 61223-3 series covers tomographic imaging including CT, MRI, and nuclear medicine. Each part specifies test conditions, measurement methods, performance indicators, and acceptance criteria tailored to the specific modality.
Tests for X-ray equipment include: radiation output reproducibility and linearity; tube voltage accuracy and radiation quality (half-value layer HVL); SID (source-to-image distance) indication accuracy; light field and radiation field coincidence; spatial resolution and contrast resolution (using line pair phantoms); dynamic range and logarithmic consistency (for DR/CR systems); and patient dose assessment (entrance surface dose ESD or dose area product DAP).
CT equipment test parameters include: CT dose index (CTDI), dose length product (DLP), CT number (Hounsfield unit) accuracy and consistency, image noise (standard deviation of water CT number), spatial resolution (MTF or line-pair method), low-contrast resolution, and slice thickness accuracy. Head and body phantoms are used for periodic calibration. Table positioning accuracy and laser alignment light accuracy are also required tests.
For nuclear medicine equipment (gamma camera, SPECT, PET), tests include: spatial resolution (point source or line source method), intrinsic uniformity (flood field image), energy calibration, sensitivity, count rate performance, and whole-body scan bed motion accuracy. For MRI equipment, tests include: signal-to-noise ratio (SNR), image uniformity, spatial linearity (geometric distortion), slice position accuracy, magnetic field homogeneity, and RF power calibration.
| Modality | Acceptance Test Items | Routine Test Frequency | Key Phantoms/Tools |
|---|---|---|---|
| General X-ray | kVp accuracy, HVL, output reproducibility, dynamic range | Monthly or quarterly | Line pair gauge, aluminum step, DAP meter |
| Fluoroscopy/Angiography | Spatial resolution, contrast, dose rate | Monthly | Low-contrast phantom, resolution grid |
| CT | CTDI, CT number accuracy, noise, MTF | Daily (air calibration) | Head/body phantoms, ACR phantom |
| MRI | SNR, uniformity, geometric distortion | Weekly | ACR phantom, Magphan phantom |
| Nuclear Medicine (SPECT/PET) | Uniformity, resolution, sensitivity | Daily (uniformity) | Co-57 flood source, point source, Jaszczak phantom |
| Mammography | AGD, contrast, resolution | Monthly | ACR mammo phantom, PMMA plates |
Healthcare facilities shall establish a comprehensive quality assurance program based on the IEC 61223 series. Program documentation should include: equipment inventory and management records, test schedules, standard operating procedures, recording forms and templates, non-conformance handling procedures, and corrective action processes. A medical physicist or trained radiologic technologist should be responsible for executing tests. Establish a test database to track performance trends over time for each equipment item.
Acceptance testing is the most critical quality control step before a new imaging device enters clinical service. A comprehensive acceptance test must be performed after installation and before the first patient examination. Results should be archived as baseline data for comparison with subsequent status and routine test results. Acceptance tests must verify all contractual performance specifications and should involve manufacturer representatives. Any non-conformances identified during acceptance testing must be resolved before clinical use begins.
All test results must be clearly documented, including: test date, operator identification, equipment identification, test conditions (kV, mA, exposure time, etc.), measured values, pass/fail determination, and remarks. Test records must be retained per regulatory requirements (typically minimum 5 years). If test results indicate non-conformance, clinical use of the equipment must be immediately suspended, and a non-conformance handling procedure initiated, including root cause analysis, corrective action (repair or calibration), and re-testing verification — with written records maintained throughout the process.
| Test Type | Purpose | Timing | Responsible Person | Documentation |
|---|---|---|---|---|
| Acceptance Test | Establish baseline performance | After installation/major repair | Medical physicist + manufacturer | Full report with contract verification |
| Status Test | Periodic performance verification | Semi-annual / annual | Medical physicist | Trend analysis report |
| Routine (Constancy) Test | Daily consistency check | Daily / weekly / monthly | Radiologic technologist | Log sheet, escalate anomalies |
| Special Test | Fault diagnosis | After repair / anomaly | Medical physicist or engineer | Repair report + verification test |
❓ How does IEC 61223 relate to other imaging quality standards and regulations?
IEC 61223 is the international standard providing the general framework for imaging equipment quality assurance. National and regional authorities have developed specific implementation requirements based on this framework, including the ACR Accreditation Program in the US, EURATOM Directive and European Guidelines in the EU, and national standards such as China’s WS 76-2020. Healthcare facilities typically combine IEC 61223 technical requirements with their national regulatory requirements to establish a compliant quality assurance program.
❓ Who should perform acceptance tests, and what qualifications are required?
Acceptance tests should be performed by a certified medical physicist or qualified imaging quality engineer. Typical qualifications include a bachelor’s or master’s degree in medical physics or biomedical engineering, plus relevant professional certification (ABMP, CCPM, or equivalent national credential). Acceptance testing involves radiation dose measurement and image performance analysis requiring in-depth understanding and practical experience. Manufacturer service personnel may assist but should not lead the testing process.
❓ What are the common causes of CT number deviation discovered during routine testing?
Common causes of Hounsfield unit deviation include: (1) X-ray tube aging causing spectral changes; (2) detector performance drift or channel failure; (3) reconstruction algorithm parameter changes; (4) phantom positioning inaccuracy or phantom degradation (liquid phantom evaporation or contamination); (5) scan room temperature variations affecting detector response. Most CT number deviations can be corrected through air and water calibration. If deviation exceeds ±5 HU after calibration, contact the manufacturer for further investigation.
❓ How should appropriate routine test frequency be determined for different imaging equipment?
Test frequency determination should be based on integrated assessment of equipment type, clinical workload, risk level, and manufacturer recommendations. High-dose equipment (CT, mammography) and critical diagnostic devices require more frequent testing. Daily tests (such as CT air calibration) should use quick, simple procedures, while monthly and quarterly tests can be more comprehensive. Initial frequencies should follow professional organization guidelines (e.g., AAPM TG-45, ACR Technical Standards), then be adjusted based on historical equipment performance stability data. Equipment with frequent stability issues should undergo increased test frequency.