Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
IEC 60731:1996/AMD1:2004 (formerly referenced in some contexts as IEC 10731-96) establishes the minimum performance requirements and test methods for dosimeters equipped with ionization chambers intended for the measurement of absorbed dose to water and dose rate in photon and electron beams used in radiotherapy. The standard applies to instruments employed for both beam calibration — typically in water phantoms at reference depth — and for quality assurance measurements in clinical settings. It covers dosimeters with coaxial cables or wireless data transmission, and includes supplementary requirements introduced by Amendment 1:2004 related to long-term stability and continuous monitoring for brachytherapy sources.
The scope explicitly excludes personal dosimeters, environmental monitors, and devices for diagnostic X‑ray fields. It focuses solely on those dosimeters that, when used with a suitable phantom, provide traceability to primary standards of absorbed dose. The standard is intended to be used by manufacturers, calibration laboratories, and radiotherapy physicists to ensure consistent, high‑quality beam measurement.
The standard mandates that dosimeters must be mechanically robust and electrically safe. They must incorporate a guarded electrometer with a leakage current not exceeding 0.5 % of the measured value for the lowest useful range, measured after a warm‑up period of 30 minutes. All connectors and switches must be clearly labeled, and the user interface must display the measured quantity (absorbed dose or dose rate) with appropriate units and decimal places.
A series of quantitative requirements define the maximum permissible deviations under reference conditions (typically 20 °C, 101.3 kPa, 50 % relative humidity). Key limits are summarized in the table below.
| Parameter | Condition / Test | Maximum Permissible Deviation |
|---|---|---|
| Linearity of response | Over the range of dose rates from 0.01 Gy/min to 10 Gy/min | ± 1.0 % of reading |
| Energy dependence (photon beams) | Reference beam quality 60Co; test at 6 MV, 10 MV, 18 MV | ± 1.5 % relative to reference |
| Energy dependence (electron beams) | Test at 6 MeV, 12 MeV, 18 MeV | ± 2.0 % relative to reference |
| Short‑term stability | 10 consecutive measurements at same dose rate | Standard deviation ≤ 0.5 % |
| Long‑term stability (AMD1:2004) | Over 8 h of continuous operation after warm‑up | Drift ≤ 1.0 % of full scale |
| Temperature coefficient | Temperature range 15 °C to 30 °C | ≤ 0.1 % per °C |
| Leakage current | With ionization chamber disconnected and input shorted | Equivalent dose rate ≤ 0.1 % of lowest measuring range |
IEC 60731:1996/AMD1:2004 does not prescribe a specific calibration method but requires that the dosimeter be calibrated in a water phantom using a secondary standard dosimeter with known calibration coefficient ND,w for the beam quality in question. The calibration must be repeated at intervals not exceeding two years, or after any repair that could affect the instrument’s response. For clinics participating in external audit programs, an annual calibration is strongly recommended.
Before a dosimeter model can be marketed as complying with IEC 60731:1996/AMD1:2004, a complete series of type tests must be performed by an accredited laboratory. These include all the performance parameters listed in the table above, as well as electrical safety tests (dielectric strength, leakage current per IEC 60601‑1), electromagnetic compatibility per IEC 61326‑1, and mechanical robustness tests (drop test, vibration).
The user is responsible for periodic routine tests to verify that the dosimeter continues to meet the standard’s specifications. Routine testing includes:
If any routine test shows a deviation outside the limits specified in the standard, the dosimeter must be withdrawn from clinical use until it is recalibrated or repaired. The standard also requires that all software used for data analysis shall be validated and version‑controlled.
To demonstrate conformity, manufacturers must issue a Declaration of Performance (DoP) listing all type‑test results and any limitations (e.g., ambient temperature range, beam quality restrictions). End‑users should retain copies of the original DoP, calibration certificates, and all daily/monthly/annual test records for at least five years.
© 2026 International Electrotechnical Commission. All rights reserved. This article is prepared for informational purposes and does not replace the official text of the standard. For certification purposes, always refer to the latest edition of IEC 60731.