Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
This standard practice provides a concise set of experiments for classifying diffusive samplers of gases and vapors, primarily for use in indoor workplace settings. It covers sampling periods ranging from 4 to 12 hours and assumes ambient wind speeds of less than 0.5 m/s. The core aim is to derive a single accuracy figure that takes into account real-world variables, including the potential for diffusive analyte loss when estimating time-weighted averages (TWAs).
Accuracy is rigorously defined in Section 3.2.2 to encompass both the imprecision of the sampler and any uncorrected bias. This definition provides a robust classification metric directly relevant to field conditions where temperature, atmospheric pressure, humidity, and wind speed are neither constant nor accurately known. The evaluation also specifically accounts for the effects of fluctuating analyte concentrations on TWAs due to diffusive mass transfer limitations.
Beyond overall accuracy, the protocol mandates testing for compliance with the manufacturer’s stated limits on capacity and requires evaluation in the presence of potential interfering compounds. This ensures the sampler’s practical applicability in complex workplace atmospheres.
| 🟦 Parameter | 📏 Specification / Requirement |
|---|---|
| 🕒 Sampling Period | 4 to 12 hours (primary scope; extendable with suitable chamber) |
| 💨 Ambient Wind Speed | < 0.5 m/s (applicable); < 0.1 m/s (excluded for static/area) |
| 🎯 Accuracy Metric | Combined imprecision and uncorrected bias (Section 3.2.2) |
| 🌡️ Field Variables | Temperature, Pressure, Humidity, Wind Speed (must be accounted for) |
| ⚡ Capacity & Interferences | Tested against manufacturer stated limits and potential interferences |
The standard also provides an efficient path for evaluating multiple analytes. Knowledge gained from similar compounds allows for data interpolation within a homologous series, drastically reducing the scope of full validation for related substances.
| 🟦 Evaluation Tier | 📐 Required Tests | 🎯 Resulting Scope |
|---|---|---|
| 🔬 Primary Analyte | Accuracy, Diffusive Loss, Capacity, Recovery, Interferences | Full Performance Classification |
| 📈 Higher MW Members | Sampling Rate, Capacity, Analytical Recovery, Interferences | Partial Validation |
The practice is designed for sampling periods of 4 to 12 hours, which is the most common range for indoor workplace exposure monitoring. It can be extended to shorter or longer periods provided a suitable exposure chamber is used to validate the extended conditions.
Accuracy is defined holistically in Section 3.2.2 to take into account both the imprecision of the sampler and any uncorrected bias. This provides a single figure of merit that is representative of expected performance under the inherently variable conditions of actual field use.
The practice explicitly does not apply to static or area sampling in wind speeds less than 0.1 m/s. Below this threshold, external diffusion outside the sampler body dominates the necessary convection from the ambient air, causing the effective sampling rate to become unreliable without a dedicated characterization chamber.
Yes. The standard recommends interpolation of data within a homologous series. If a primary compound is fully evaluated (accuracy, diffusive loss, capacity), higher molecular weight members in the series can receive a partial validation focusing on sampling rate, capacity, analytical recovery, and interferences. The expensive diffusive loss test can be omitted if proven negligible for the series.