Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D6238-98 (Reapproved 2024) defines a standard procedure for measuring Total Oxygen Demand (TOD) in water samples. It is a rapid combustion method suitable for a wide variety of aqueous matrices, from clean waters to industrial brines, and is adaptable for both laboratory and on-stream process monitoring.
This test method is designed to cover the determination of total oxygen demand within a standard range of 100 mg/L to 100,000 mg/L. The method relies on a carrier gas with a known oxygen concentration; the standard specifies that the oxygen level in the carrier gas must be set to two to four times the maximum expected oxygen demand of the sample to ensure optimal sensitivity and accuracy. The analysis is based on the change in oxygen reading of the carrier gas compared to that when a sample is introduced into the reaction zone.
| 🟦 📏 Parameter | 🎯 Specification |
|---|---|
| Standard Analysis Range | 100 mg/L to 100,000 mg/L |
| Minimum Recommended Range | 2 mg/L to 100 mg/L |
| Theoretical Upper Limit | 250,000 mg/L to 500,000 mg/L |
| Carrier Gas O₂ Requirement | 2 to 4 times max expected O₂ demand |
The operational boundaries of this test method are tightly controlled by the detection system. The lower measurement range is dictated by the stability of the baseline oxygen detector output, which is a function of the permeation system temperature, carrier gas flow rate, oxygen detector temperature, and reference sensor voltage. Combined, these variables limit the minimum recommended range to 2 mg/L to 100 mg/L.
The upper measurement range is physically limited by the maximum oxygen concentration in the carrier gas (100%). Under the recommended condition where the carrier gas concentration is two to four times the desired oxygen demand, this limits the maximum possible to between 250,000 mg/L and 500,000 mg/L. The method is applicable to all oxygen-demanding substances that can be physically injected, though the injector opening limits the maximum particle size. For water-insoluble liquids or solids, preliminary sample treatment methods are described in Annex A2 of the standard.
| 🎯 ⚡ Operational Parameter | 📐 Requirement / Limit |
|---|---|
| Lower Range Stability Factors | Temperature, flow rate, sensor voltage |
| Upper Carrier Gas O₂ | 100% (maximum physical concentration) |
| Sample Injection Limit | Particle size limited by injector opening |
| Solids Pretreatment | Refer to Annex A2 for guidance |
This test method is particularly well-suited for measuring oxygen demand in industrial effluents and process streams, including on-stream monitoring applications. The standard explicitly defines critical limitations for other matrices:
The standard test method covers the determination of total oxygen demand in the range from 100 mg/L to 100,000 mg/L in water and wastewater, including brackish waters and brines. Larger concentrations may be determined by suitable dilution of the sample.
Under recommended conditions, the carrier gas concentration should be between two to four times the maximum desired oxygen demand of the sample. This ensures the change in oxygen reading is accurately detectable by the oxygen detector.
No. The standard explicitly states that its use for monitoring natural waters is greatly limited by interferences defined in the standard text. Furthermore, its application for monitoring secondary sewage effluents is not established.
The maximum size of particles that can be analyzed is directly limited by the injector opening. If oxygen-demanding substances that are water-insoluble liquids or solids are present, a preliminary treatment method, as described in Annex A2, may be required to ensure proper injection and combustion.