Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D5955-02 (Reapproved 2017) defines standard test methods for the sampling and analysis of respirable suspended particles (RSP) and the estimation of the RSP fraction attributable to environmental tobacco smoke (ETS). This method, based on the measurement of ultraviolet absorbance (UVPM) and fluorescence (FPM), provides critical tools for indoor air quality assessments.
These test methods cover the collection of total RSP on a preweighed, 1.0‑µm pore size polytetrafluoroethylene (PTFE) membrane filter. The collected material is extracted in methanol, and the extract is analyzed for total ultraviolet absorbance or fluorescence to estimate the ETS contribution. The methods are compatible with the solanesol‑based approach (ASTM D6271) but do not require it. Sampling components include a filter cassette connected to a particle size separating device on the inlet end and a sampling pump on the outlet end, making it suitable for both personal and area sampling.
The procedure requires careful flow rate adjustment per Practice D5337. The sample duration is limited entirely by the filter’s mass capacity, which is approximately 2000 µg. The standard has been evaluated for sample durations ranging from a minimum of 1 hour up to a maximum of 24 hours. All measurements and calculations strictly adhere to SI units as the standard.
The following tables summarize the critical specifications and detection limits defined by the standard at a sampling rate of 2 L/min.
| 🟦 Parameter | 📐 Specification |
|---|---|
| Filter Media | Polytetrafluoroethylene (PTFE) Membrane |
| Pore Size | 1.0 µm |
| Maximum Filter Loading | ~2000 µg |
| Sample Duration Range | 1 to 24 hours |
| Standard Sampling Rate | 2 L/min |
| 🟦 Method | 📏 Duration | 🎯 LOD (µg/m³) | ⚡ LOQ (µg/m³) |
|---|---|---|---|
| UVPM | 1 hour | 2.5 | 8.3 |
| UVPM | 8 hours | 0.3 | 1.0 |
| FPM | 1 hour | 1.4 | 4.7 |
| FPM | 8 hours | 0.2 | 0.6 |
These test methods are fully compatible with but do not require the determination of solanesol (D6271). They offer alternative approaches (UVPM and FPM) for estimating the ETS contribution to RSP and can be used to provide a more comprehensive analysis of indoor air particulate matter.
The standard mandates the use of a preweighed, 1.0‑µm pore size polytetrafluoroethylene (PTFE) membrane filter housed in a standard filter cassette assembly. The filter must be handled with care to avoid contamination or damage prior to sampling.
For an 8-hour sample at 2 L/min, the FPM limit of detection (LOD) is 0.2 µg/m³ and the limit of quantitation (LOQ) is 0.6 µg/m³. For a 1-hour sample, the LOD is 1.4 µg/m³ and the LOQ is 4.7 µg/m³.
The sample duration is limited by the capacity of the membrane filter, which is approximately 2000 µg. The standard has been validated for sample durations from 1 hour up to a maximum of 24 hours under typical indoor concentration levels.