Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
The ASTM D5812-96 (Reapproved 2002) standard, an American National Standard, defines the test method for the determination of organochlorine pesticides in water by capillary column gas chromatography. This rigorous method requires significant analyst expertise and emphasizes the importance of proper chromatographic separation and result confirmation techniques.
This test method covers the capillary gas chromatographic determination of various organochlorine pesticides, including their degradation products and related compounds in finished drinking water. While primarily validated for drinking water, its applicability to other aqueous matrices must be determined by the user. The standard explicitly lists a wide range of analytes including Aldrin, Chlordane isomers, DDT and its metabolites, Dieldrin, Endrin, Heptachlor, Hexachlorobenzene, and Methoxychlor. The extract derived from this procedure may be analyzed using the GC conditions prescribed in Test Method D 5175 (capillary column).
| 🧪 Pesticide / Compound | 🔢 CAS Registry Number |
|---|---|
| Aldrin | 309-00-2 |
| α-BHC | 319-84-6 |
| β-BHC | 319-85-7 |
| δ-BHC | 58-89-9 |
| α-Chlordane | 5103-71-9 |
| γ-Chlordane | 5103-74-2 |
| 4,4′-DDD | 72-54-8 |
| 4,4′-DDE | 72-55-9 |
| 4,4′-DDT | 50-29-3 |
| Dieldrin | 60-57-1 |
| Endosulfan sulfate | 1031-0708 |
| Endrin | 72-20-8 |
| Heptachlor | 76-44-8 |
| Heptachlor epoxide | 1024-57-3 |
| Methoxychlor | 72-43-5 |
This test method is restricted to use by or under the supervision of analysts experienced in the use of GC and interpretation of gas chromatograms. Table 1 and Table 2 of the standard list the applicable concentration ranges and the precision and bias statements. Each analyst must demonstrate the ability to generate acceptable results using the procedures described in Section 12. Values are stated in SI units, which are to be regarded as the standard.
| 📐 Procedural Requirement | 📏 Specification |
|---|---|
| Analyst Qualification | Experienced in GC usage and chromatogram interpretation (Section 1.4) |
| Performance Validation | Proficiency must be demonstrated per Section 12 |
| Identification Confirmation | Confirmation by at least one additional technique required for unfamiliar samples |
| Reference Method | GC conditions from Test Method D 5175 (capillary column) |
When this test method is used to analyze unfamiliar samples for any or all of the listed analytes, analyte identifications and concentrations should be confirmed by at least one additional technique. The rigorous confirmation requirements ensure accuracy in complex matrices where interferences can lead to false positives or misidentification. The method development data focused on single analytes, and users must exercise caution if extending the method to multi-congener constituents.
🔍 What specific water types are covered by this standard?
The test method is specifically validated for finished drinking water. Section 1.1 clearly states that its applicability to other aqueous matrices, such as surface water or wastewater, must be determined by the user.
💡 What happens if two target compounds co-elute on the GC column?
According to Section 1.5, analytes that are not separated by either the primary or secondary column cannot be identified and measured individually unless an alternative technique exists for their identification and quantitation. The analyst must resolve the co-elution or use a confirmatory method.
⚡ Who is qualified to perform this complex GC analysis?
Section 1.4 restricts this test method to use by or under the supervision of analysts experienced in the use of gas chromatography and the interpretation of gas chromatograms. Each analyst must also pass proficiency testing per Section 12.
📌 Why is an additional confirmation technique so strongly recommended?
Section 1.6 states that when unfamiliar samples are analyzed, identifications and concentrations should be confirmed by at least one additional technique. This is critical for environmental samples where unique degradation products or matrix interferences can easily lead to misidentification of target analytes.