Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D5391-23 is the definitive standard for measuring the electrical conductivity and resistivity of high purity water. This standard is specifically applicable to water samples with a conductivity below 10 µS/cm (equivalent to a resistivity above 0.1 MΩ·cm). A critical and mandatory requirement specified in the scope is that the water sample must be continuously flowing during measurement. This test method is suitable for both continuous on-line monitoring directly in process lines and periodic measurements using side stream sample lines, accommodating high temperature or high pressure sample streams. The standard mandates the use of SI units for all reported values.
The core methodology described in the standard involves a conductivity or resistivity measurement cell equipped with an integral temperature sensor or compensator. This assembly is installed in a flowing, closed system to isolate the high purity water sample. Preventing contact with the atmosphere and carefully managing the wetted materials are critical steps to avoid trace contamination from sources like carbon dioxide absorption, which would rapidly increase the conductivity and invalidate the measurement.
A key technical definition found in Section 3 of the standard is the cell constant (n). It is defined as the ratio of the length of the electrical path between the electrodes (L) to the cross-sectional area of the solution (A), expressed in units of reciprocal centimeters (cm⁻¹). Selecting the correct cell constant is paramount for accuracy. For the low ionic concentrations found in high purity water, cell constants must be low to minimize polarization and stray electrical interference.
| 🟦 Application | 📏 / 📐 Cell Constant (n) | ⚡ Conductivity / Resistivity Range |
|---|---|---|
| High Purity Water (D5391-23) | 0.001 – 0.1 cm⁻¹ | < 10 µS/cm (> 0.1 MΩ·cm) |
| Standard Reference Cell | 1.0 cm⁻¹ | > 10 µS/cm (< 0.1 MΩ·cm) |
The standard explicitly states that static grab sampling cannot be used for high purity water. Even brief exposure to the atmosphere or a sample container will introduce contaminants and dissolved gases, rendering the conductivity or resistivity measurement invalid. A continuous, flowing stream in a closed system is mandatory for representative sampling.
Using a standard cell constant of 1.0 cm⁻¹ for high purity water (< 10 µS/cm) can cause significant polarization errors. The standard mandates lower cell constants (0.001 to 0.1 cm⁻¹) which are specifically matched with direct-reading instrument ranges to ensure accurate measurement of extremely low ionic concentrations.
The standard covers high purity water samples with an electrical conductivity below 10 µS/cm, which corresponds to an electrical resistivity above 0.1 MΩ·cm.
High purity water is extremely aggressive in its solvent properties. If left static, it rapidly absorbs atmospheric carbon dioxide and leaches ions from container walls, drastically altering its conductivity. A flowing, closed system prevents this contamination and ensures the measurement reflects the true process water quality.
The cell constant (n) is the ratio of the electrode path length to the cross-sectional area of the solution (n = L/A), with units of cm⁻¹. For high purity water measurements under this standard, the cell constant is typically between 0.001 and 0.1 cm⁻¹ to prevent electrical interference.
The standard references several others, including D1125 (Test Methods for Conductivity and Resistivity), D1129 (Terminology Relating to Water), and D3864 (Guide for On-Line Monitoring Systems for Water Analysis).