Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D5197-21 establishes a standardized procedure for the determination of formaldehyde and other carbonyl compounds in air using active sampler methodology. This method is widely applied in indoor air quality investigations, industrial hygiene assessments, and environmental monitoring, functioning as a comprehensive modification of U.S. EPA Method TO-11A.
This test method is based on the reaction of carbonyl compounds with acidified 2,4-dinitrophenylhydrazine (DNPH) to form stable hydrazone derivatives. Air is actively drawn through a cartridge containing chromatographic-grade silica gel coated with DNPH. Following sampling, the derivatives are extracted and analyzed using High Performance Liquid Chromatography (HPLC) or Ultra-High Performance Liquid Chromatography (UHPLC) with UV absorption or photodiode array (PDA) detection.
The sampling method yields a time-weighted average (TWA) concentration. It is validated for long-term sampling (1 to 24 hours) and short-term sampling (5 to 60 minutes) of formaldehyde. The user-prepared silica gel cartridges have a particle size of 55 to 105 µm which results in a high pressure drop, specifically greater than 8 kPa at 1.0 L/min. This physical limitation restricts the use of low-pressure battery-powered personal sampling pumps.
| 🟦 Parameter | 📏 Specification / Value |
|---|---|
| Validated Sampling Flow Rate | Up to 1.5 L/min |
| Cartridge Pressure Drop | > 8 kPa at 1.0 L/min |
| Sampling Duration Range | 5 min to 24 h |
| Sampler Type | Active, DNPH-coated silica gel |
| Detection System | HPLC / UHPLC with UV or PDA |
Shorter sampling durations or lower flow rates increase method detection limits and may introduce greater variability between co-located samplers. Project-specific data quality objectives must dictate the chosen flow rate and duration. For optimal quantification of formaldehyde at 10 ppbv, a 1-hour sample at 1 L/min is typically required.
The standard has been successfully applied to a wide suite of carbonyl compounds. The data obtained provides total concentrations from which TWA concentrations can be calculated. The working concentration range for a 1-hour sample at 1 L/min spans from approximately 10 ppbv/v (12 µg/m³) to 1 ppmv/v (1.2 mg/m³).
| 🎯 Compound | ⚡ Type | 📐 Typical Source / Application |
|---|---|---|
| Formaldehyde (HCHO) | Aldehyde | Building materials, indoor emissions |
| Acetaldehyde | Aldehyde | Combustion products, industrial emissions |
| Acetone | Ketone | Industrial solvents, general indoor air |
| 2-Butanone (MEK) | Ketone | Coatings, adhesives, industrial solvents |
| Benzaldehyde | Aldehyde | Aromatic emissions, chemical intermediates |
| Hexanal | Aldehyde | Lipid oxidation, food processing environments |
Other successfully quantified carbonyls include propanal, butyraldehyde, isovaleraldehyde, valeraldehyde, o/m/p-tolualdehyde, and 2,5-dimethylbenzaldehyde.
While sharing core DNPH chemistry, D5197-21 expands the scope beyond TO-11A to include a broader range of ketones and aldehydes, such as 2,5-dimethylbenzaldehyde and hexanal. It also provides explicit guidance for both traditional HPLC and modern UHPLC instrumentation.
For a 1-hour sample at 1 L/min, the method reliably quantifies formaldehyde and other carbonyls in the range of approximately 10 ppbv/v (12 µg/m³) up to 1 ppmv/v (1.2 mg/m³). Lower limits can be achieved by increasing the total sampling volume and implementing strict contamination controls.
Generally, no. The specific silica gel cartridges used in this method have a high resistance to flow (pressure drop >8 kPa). This limits their compatibility with the low-pressure, battery-powered pumps typically used for personal sampling. It is best suited for area or stationary source sampling.
After sampling, the cartridges are eluted with acetonitrile to recover the hydrazone derivatives. The solution is then injected into an HPLC or UHPLC system. Compounds are separated on a C18 column and detected using UV absorption (typically at 360 nm) or a photodiode array detector for qualitative and quantitative analysis.