Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D5156-22 specifies a continuous measurement method for ozone (O₃) concentrations using ultraviolet (UV) absorption. This test method is validated for use across ambient, workplace, and indoor atmospheric environments. The officially recognized concentration range is 10 to 2000 µg/m³ of O₃ in air, which is equivalent to 5 ppb(v) to 1 ppm(v).
The operating principle relies on the strong absorption of ultraviolet light by ozone molecules at a wavelength of 253.7 nm. The instrument measures the attenuation of a UV beam through a sample cell, and the ozone concentration is calculated from the resulting absorbance using the well-established absorption coefficient for ozone at this wavelength.
Calibration traceability is a critical aspect of the standard. An Absolute Ultraviolet Photometer serves as the primary standard. By definition, this is a photometer whose design and maintenance allow it to measure the absorbance of O₃ mixtures without external absorption standards. All field and transfer standard instruments must be directly compared against such a primary standard, following the rigorous procedures outlined in Practices D5011 and D5110.
| 🟦 Specification Category | 📏 Parameter / Value |
|---|---|
| Analyte | Ozone (O₃) |
| Measurement Method | Ultraviolet (UV) Absorption |
| Wavelength of Operation | 253.7 nm |
| Minimum Concentration (Range) | 10 µg/m³ (5 ppb(v)) |
| Maximum Concentration (Range) | 2000 µg/m³ (1 ppm(v)) |
| Primary Calibration Standard | Absolute Ultraviolet Photometer |
The method operates on the principle of UV absorption. Ozone molecules absorb light at 253.7 nm. The photometer measures the intensity of light before and after passing through the sample; the difference in intensity is directly proportional to the ozone concentration as defined by the Beer-Lambert law.
The standard is validated for the continuous measurement of ozone in the range of 10 to 2000 µg/m³, which corresponds to 5 ppb(v) to 1 ppm(v) in air. Measurements outside this range are not covered under this specific standard.
Calibration is traceable to a Primary Standard, specifically the Absolute Ultraviolet Photometer. This device calculates O₃ concentrations directly from its physical design and the known absorption coefficient, without needing external physical standards. Practices D5011 and D5110 detail the specific procedures for implementing this traceability chain.
As outlined in Section 6, the primary interferences are chemical species that also absorb UV light at 253.7 nm. The most significant are aromatic volatile organic compounds (VOCs) like benzene and toluene, molecular mercury vapor, and high concentrations of nitrogen dioxide (NO₂). The standard mandates careful evaluation of the air matrix for these compounds prior to deployment.