Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D5151-19 (Reapproved 2023) is the standard test method for detecting holes in medical gloves, covering both surgical and examination glove types. This globally recognized procedure (developed under WTO TBT principles) serves as a critical quality control measure to identify gloves compromised by water-leaking holes, ensuring they do not reach clinical use.
This reference method is designed for evaluating glove samples and relies entirely on a simple visual water leak test. It is the responsibility of the user to apply proper safety, health, and regulatory limits as outlined in Section 1.6.
The procedure utilizes a clear plastic mandrel (50 mm schedule 80 PVC pipe, 38 cm ± 1 cm long, 6.0 cm ± 0.5 cm outside diameter) as a support vessel for the glove. The glove cuff is mounted over the bottom of the mandrel with a maximum of 40 mm of the cuff secured using a clamp or rubber band. A minimum of 1000 cm³ of room temperature water (20°C to 30°C) from a graduated cylinder or beaker is poured into the glove through the mandrel. The exterior of the glove must remain completely dry throughout the setup phase.
| 🟦 Parameter | 📐 Specification |
|---|---|
| Mandrel Outside Diameter | 6.0 ± 0.5 cm |
| Mandrel Length | 38 ± 1 cm |
| Minimum Water Volume | 1000 cm³ |
| Water Temperature | 20°C to 30°C |
| Glove Cuff Overlap | ≤ 40 mm |
| Inspection Period | Immediate + after 2 minutes |
Results are definitive according to Section 7. Any glove showing a droplet, stream, or any other type of water leakage on the body of the glove is considered to have failed the test. However, water leakage occurring solely above the junction of the glove cuff and mandrel is not considered a test failure. If this specific leak occurs, the operator must remove the glove, dry the outside, and retest it following the standard procedure.
| 🎯 Observation | ⚡ Result (Section 7) |
|---|---|
| No water leakage anywhere on the glove | Pass |
| Droplet or stream on the glove body | Failure |
| Leakage solely at the cuff/mandrel junction | Retest (after drying exterior) |
The standard explicitly states in Section 1.3 that the smallest hole size that will allow water leakage has not been determined and is beyond the scope of this test method. Detection depends on water pressure, surface tension, and exposure time.
No. As noted in Section 1.2, this test is strictly limited to detecting holes that allow water leakage under these specific conditions. It cannot guarantee the absence of microscopic holes or defects below the physical detection threshold of the water leak test.
Per Section 6.4, the operator must visually inspect the glove for immediate water leakage right after filling. The glove is then allowed to hang for 2 minutes. After this period, a second visual inspection for water leakage is conducted to determine the final result.
According to Section 7.2, water leakage above the junction of the glove cuff and mandrel is not a test failure. The operator should remove the glove, thoroughly dry the outside, and retest the same glove to confirm whether the body of the glove is actually compromised.