ASTM D4420 – 94 (Reapproved 1999) defines a robust gas chromatographic methodology for quantifying aromatic hydrocarbons in finished motor gasoline. The standard provides validated procedures for determining benzene, toluene, C8 aromatics, C9 and heavier aromatics, and total aromatics, yielding essential data for quality control and health exposure assessments.
The procedure employs a sophisticated two-column chromatographic system paired with a thermal conductivity detector (TCD). A precise sample volume is injected into a polar column, where non-aromatic hydrocarbons are rapidly eluted and directed to the detector reference side before being vented. Immediately prior to the elution of benzene, the polar column is backflushed, sweeping the aromatic fraction onto a secondary nonpolar column. Aromatics separate according to boiling point on this analytical column. Once the C8 aromatics have fully eluted, flow through the nonpolar column is reversed to backflush the C9 and heavier aromatics through the detector. Quantitation is achieved by comparing integrated peak areas against response factors derived from a gravimetrically prepared standard blend of known aromatic composition.
The chromatograph must be configured with a direct vaporizing injector, a precise backflush system (utilizing an eight-port rotary valve or an eleven-port linear valve), and a thermal conductivity detector—which can be either a single dual-filament unit or two single-filament detectors. The method is validated across specific concentration intervals for optimal performance.
| 🟦 Aromatic Component | 📏 Validated Range (Liquid Vol. %) |
|---|---|
| Benzene | 0.1 – 5 |
| Toluene | 1 – 20 |
| C8 Aromatics | 3 – 25 |
| C9 and Heavier Aromatics | 5 – 30 |
| Total Aromatics | 10 – 80 |
| 🛠️ Apparatus Component | ⚡ Required Specification |
|---|---|
| Injector | Direct Vaporizing |
| Backflush System | 8-Port Rotary or 11-Port Linear Valve |
| Detector | Thermal Conductivity (Dual or Single-Filament) |
| Column 1 | Polar Liquid Phase |
| Column 2 | Nonpolar Liquid Phase |
Knowledge of the aromatic profile in gasoline is indispensable for assessing product quality and evaluating the potential health hazard for personnel involved in handling and using gasoline. All results are reported to the nearest 0.1 % by liquid volume, with SI units serving as the standard application.
A significant caveat for the user is that the standard’s applicability to gasoline containing oxygenates (e.g., ethers and alcohols) has not been determined. Analytical chemists are also directed to Test Method D 3606 exclusively for benzene and toluene analysis.
The method is designed to quantify Benzene, Toluene, C8 aromatics (xylenes, ethylbenzene), C9 and heavier aromatics, and the cumulative Total Aromatics in finished motor gasoline.
Non-aromatics elute first from the polar column and are vented. Aromatics are then backflushed onto the nonpolar column for boiling point separation. After C8 elutes, C9+ components are backflushed through the detector, ensuring complete analysis without excessively long run times.
The standard validates specific ranges: Benzene (0.1–5 %), Toluene (1–20 %), C8 (3–25 %), and C9+ (5–30 %). The total aromatics range is validated from 10 % to 80 % by liquid volume.
Yes. According to Section 1.1 of the standard, the test method is applicable to both finished motor gasoline and gasoline blending components, providing flexibility across the fuel production and blending workflow.