Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D3797-05 specifies a standard gas chromatographic test method for analyzing the purity of o-xylene and quantifying its normally occurring impurities. The impurities covered in this test method include nonaromatic hydrocarbons, benzene, toluene, p– and m-xylenes, cumene, styrene, and ethylbenzene.
The method is specifically applicable for impurity concentrations ranging from 0.001 to 2.000 % and is designed for o-xylene purities of 98 % or higher. For purposes of determining conformance with this standard, all observed or calculated values must be rounded to the nearest unit in the last right-hand digit of the specification limit, strictly following the rounding-off method described in Practice E 29.
| 🟦 Analyte / Property | 📏 Specified Range |
|---|---|
| o-Xylene Purity | ≥ 98.0 % |
| Nonaromatic Hydrocarbons | 0.001 – 2.000 % |
| Benzene | 0.001 – 2.000 % |
| Toluene | 0.001 – 2.000 % |
| p– and m-Xylenes | 0.001 – 2.000 % |
| Ethylbenzene | 0.001 – 2.000 % |
| Cumene (Isopropylbenzene) | 0.001 – 2.000 % |
| Styrene | 0.001 – 2.000 % |
A known amount of an appropriate internal standard is added to the sample. The analysis is performed using a gas chromatograph equipped with a flame ionization detector (FID) and a polar fused-silica capillary column. The apparatus must include a splitter injector suitable for use with the capillary column and must provide sufficient sensitivity to obtain a minimum peak response for the lowest required impurity levels.
| 🎯 Component | ⚡ Requirement |
|---|---|
| Gas Chromatograph | High resolution capability with FID |
| Column Type | Polar Fused-Silica Capillary |
| Injector | Splitter Injector |
| Detector | Flame Ionization Detector (FID) |
Impurity concentrations are measured relative to the internal standard peak areas. The installation of the fused-silica capillary column should be performed in accordance with Practice E 1510.
To calculate the o-xylene purity, the sum of the concentrations of all determined impurities is subtracted from 100.00 %. As stated in Section 5.2 of the standard, absolute purity cannot be determined if unknown impurities are present; the reported purity is derived strictly from the expected impurities listed in the scope.
When reporting results to determine conformance with specifications, all observed or calculated values must be rounded to the nearest unit in the last right-hand digit of the specification limit. This rounding must be performed in strict accordance with the procedures outlined in Practice E 29, as required by Section 1.3.
The test method is designed to quantify nonaromatic hydrocarbons, benzene, toluene, p– and m-xylenes, cumene, styrene, and ethylbenzene.
Purity is determined by subtracting the sum of all determined impurity concentrations from 100.00 %.
This test method is applicable for individual impurities at concentrations from 0.001 to 2.000 %.
Section 1.3 of the standard mandates that all values be rounded using the rounding-off method of Practice E 29.