D3447-01 – Standard Test Method Technical Guide

🔬 Purity of Halogenated Organic Solvents – Standard Test Method D3447-01

This standard, under ASTM Committee D26, describes determining common impurities in halogenated organic solvents. It uses a temperature-programmed gas chromatographic technique with a TCD or FID, achieving sensitivity in the ppm range.

🧪 Principle and Apparatus Setup

The method separates the solvent from its impurities via a packed column containing 30 % silicone fluid on 80/100 mesh diatomaceous earth or a similar capillary column.

🟦 Component📏 Specification
Gas Chromatograph1-mV recorder/integrator, TCD or FID
Column6-m x 3.2-mm OD SS packed with 30% silicone fluid on 80/100 mesh diatomaceous earth
Carrier GasHelium, Research Grade
Syringe10-µL GC syringe

⚙️ Procedure and Chromatographic Conditions

Install the column, set He flow to 40 mL/min, and configure the GC as follows:

<

📥 Standard Documents Download

🔒
Please wait 10 seconds, the download links will appear after the ad loads

Leave a Reply

Your email address will not be published. Required fields are marked *

🎯 Parameter⚡ Value
Injection Port Temp200°C
Detector Temp220°C
Oven Init / Final Temp50°C / 170°C
Program Rate10°C/min
Carrier Flow