Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D3362-05 is a standardized gas chromatographic test method for determining the purity of commercially available acrylate esters. It specifically covers methyl acrylate, ethyl acrylate, n-butyl acrylate, and 2-ethylhexyl acrylate. The method simultaneously quantifies common impurities such as alcohols and other esters, ethers, and several unidentified compounds.
To ensure a complete mass balance, the standard dictates that water and acidity must be measured by other appropriate ASTM procedures (Test Methods D1364 and D1613). The results from these tests are used to normalize the chromatographic values, providing a rigorous and accurate assessment of product purity.
| 🟦 Analyte | 📏 Impurities Detected by GC | 🎯 Normalization Factors |
|---|---|---|
| Methyl Acrylate | Alcohols, other esters, ethers | Water & Acidity (D1364, D1613) |
| Ethyl Acrylate | Alcohols, other esters, ethers | Water & Acidity (D1364, D1613) |
| n-Butyl Acrylate | Alcohols, other esters, ethers | Water & Acidity (D1364, D1613) |
| 2-Ethylhexyl Acrylate | Alcohols, other esters, ethers | Water & Acidity (D1364, D1613) |
The method summary involves introducing a representative specimen into a gas chromatographic column. An inert carrier gas transports the components, separating the acrylate ester from its impurities. The separated components are measured by a detector and recorded as a chromatogram.
The standard requires a gas chromatograph equipped with either a thermal conductivity detector (TCD) or a flame ionization detector (FID). The system must demonstrate sufficient sensitivity and stability to obtain for 0.01 weight % of impurity a recorder deflection of at least 2 mm at a signal-to-noise ratio of at least 5 to 1. The column specified is 6 m (20 ft) in length.
| 🟦 Parameter | 📐 Required Specification |
|---|---|
| Detector Options | Thermal Conductivity (TCD) or Flame Ionization (FID) |
| Column Length | 6 m (20 ft) |
| Minimum Detectable Impurity | 0.01 weight % |
| Signal-to-Noise Ratio at Threshold | 5:1 (minimum) |
| Recorder Deflection at Threshold | 2 mm (minimum) |
The raw chromatogram is interpreted by applying component attenuation and detector response factors to the peak areas. The relative concentrations of the acrylate ester and its impurities are determined by relating the individual peak responses to the total peak response.
Water content is quantified using Fischer Reagent Titration (Test Method D1364) and acidity is measured per Test Method D1613. These values are then integrated into the GC data to normalize the overall composition. For specification purposes, the sum of all impurities (including water and acidity) can be subtracted from 100 to report the absolute assay of the acrylate ester.
🔍 Which acrylate esters are covered by this test method?
The standard applies to methyl acrylate, ethyl acrylate, n-butyl acrylate, and 2-ethylhexyl acrylate.
💡 How is the final purity (assay) reported?
The purity is typically determined by totaling all impurities—those detected by GC, plus water and acidity measured by wet chemistry—and subtracting this sum from 100%.
⚡ What detector types are acceptable for this test