Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
The ASTM D3316-06 (Reapproved 2022) standard defines a rigorous method for evaluating the corrosiveness of perchloroethylene to copper metal. This accelerated stability test is critical for industries such as dry-cleaning and metal degreasing, serving as a primary guide for qualifying solvent grades. The method exposes meticulously prepared copper strips to both the liquid and vapor phases of boiling perchloroethylene for a duration of 72 hours under the constant illumination of a 100 W light bulb. The test directly simulates the demanding conditions that solvents face in service, where thermal and UV stress can accelerate the formation of corrosive byproducts.
Values stated in SI units are regarded as standard. Users of this standard must establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
The test relies on a precision glassware assembly. The apparatus consists of an Allihn bulb condenser (34/45 standard-taper bottom joint), a Soxhlet extractor (30 mm inside diameter, 24/40 bottom joint, 34/45 upper joint), and a 300 mL flask (24/40 standard-taper joint) with a variable control heater. Copper test specimens are prepared from 0.05 mm thick electrolytic foil, cut into three strips measuring 20 mm by 75 mm. These strips are scribed with an identification number, immersed in concentrated hydrochloric acid until uniformly bright and free of tarnish, rinsed thoroughly in running water, dried, and weighed to the nearest 0.1 mg on an analytical balance.
| 🛠️ Component | 📐 Specification | 📏 Critical Detail |
|---|---|---|
| Boiling Flask | 300 mL | 24/40 Standard-Taper Joint |
| Soxhlet Extractor | 30 mm Inside Diameter | 24/40 Bottom / 34/45 Upper Joint |
| Condenser | Allihn Bulb Type | 34/45 Bottom Joint |
| Copper Foil/Specimens | 20 mm x 75 mm Strips | Electrolytic Foil, 0.05 mm Thick |
| Light Source | 100 W Incandescent Bulb | Continuous exposure during test |
Stability is assessed by two primary metrics after the 72-hour boiling and condensation period. Firstly, the weight loss of the copper strips is measured to quantify direct corrosive attack. Secondly, the presence of acid formation in the perchloroethylene is evaluated, which indicates chemical decomposition of the solvent itself. Together, these data points definitively characterize the solvent’s stability and corrosivity. Solutions prepared for the acid formation analysis should comply with practices outlined in ASTM E200 (Practice for Preparation, Standardization, and Storage of Standard and Reagent Solutions).
| 📏 Parameter | 📊 Condition / Metric | 🎯 Purpose |
|---|---|---|
| Exposure Duration | 72 Hours | Standard accelerated aging period |
| Illumination | 100 W Light Bulb | Simulates UV/Thermal stress |
| Specimen Quantity | Three Copper Strips | Provides replicate data |
| Primary Metrics | Cu Weight Loss & Acid Formation | Determines grade acceptability |
The test evaluates the corrosive stability of perchloroethylene by exposing clean copper strips to both the liquid and vapor phases of the boiling solvent for 72 hours under constant light. Degradation of the solvent creates acidic byproducts that attack the copper, quantifiable as weight loss and acid formation.
Three copper strips are required per test. They must be cut from 0.05 mm thick electrolytic copper foil and measure precisely 20 mm by 75 mm. They are weighed to the nearest 0.1 mg before and after the test.
The boiler flask requires a 24/40 standard-taper joint. The Soxhlet extractor has a 24/40 bottom joint and a 34/45 upper joint. The Allihn condenser has a 34/45 bottom joint. Using the correct joints ensures a vapor-tight system.
Different failure mechanisms can occur in the liquid phase versus the vapor phase. As the perchloroethylene boils and condenses in the Soxhlet apparatus, the copper strip is alternately bathed in hot liquid and exposed to hot vapor, simulating the most severe conditions found in industrial degreasers and dry-cleaning recovery tumblers.