Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D3271-87 (Reapproved 2012) outlines a standardized technique for directly injecting solvent-reducible paint samples into a gas chromatograph to obtain a chromatogram for solvent composition analysis. This practice is intended for qualitative assessment only and is not designed for quantitative measurements. All values are stated in SI units, and users are responsible for establishing appropriate safety and health practices before use.
The gas chromatograph must be capable of temperature programming and equipped with either a thermal conductivity detector (TCD) or a flame ionization detector (FID). A recorder with a full-scale response time of 2 seconds or less and maximum noise of 0.03% of full scale is required. The column, which can be packed or capillary, must achieve a resolution such that the depression between two adjacent peaks does not exceed 50% of the smaller peak.
| 🟦 Column Material | 📏 Specification | 🎯 Mesh Size |
|---|---|---|
| Polyethylene glycol (MW 20,000) + Diisodecyl phthalate | Liquid phases on diatomaceous earth | 60–80 mesh (250–175 µm) |
| Ethylvinylbenzene/Divinylbenzene copolymer | Porous beads | 60–80 mesh (250–177 µm) |
Carrier gases are selected based on the detector type: helium or hydrogen for TCD, and nitrogen, helium, or argon for FID. Column packing materials as listed above have been used successfully. Liquid charging devices such as micro syringes are required, with disposable types preferred to avoid contamination.
| 🔍 Detector Type | ⚡ Recommended Carrier Gases |
|---|---|
| Thermal Conductivity (TCD) | Helium, Hydrogen |
| Flame Ionization (FID) | Nitrogen, Helium, Argon |
The practice describes techniques for injecting whole paint samples into a gas chromatograph to separate and analyze solvent composition qualitatively.
No, it is explicitly stated that this practice is not designed to be quantitative.
The standard recommends using either a thermal conductivity detector (TCD) or a flame ionization detector (FID) as per the apparatus requirements.
When hydrogen is used, special precautions are required to prevent gas leakage and potential explosions. Users must follow safety guidelines.