Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D3104-14a (Reapproved 2018) defines the Mettler Softening Point Method for determining the softening point of pitches with softening points in the range of 50°C to 180°C. This test method yields results comparable to Test Method D2319 (Cube-in-Air Method) for temperatures above 80°C (176°F). The values stated in SI units are regarded as standard; values in parentheses are for information only.
Pitch does not undergo a solid-liquid phase change when heated, and thus does not have a true melting point. As temperature increases, pitch gradually softens or becomes less viscous. Consequently, the softening point must be determined by an arbitrary yet precisely defined method to ensure reproducible results. This test method is useful for determining the consistency of pitches as one element in establishing the uniformity of shipments or sources of supply. The standard was developed in accordance with internationally recognized principles on standardization.
This test method references several ASTM standards, including A314 for stainless steel billets, D2319 for the cube-in-air method, and D4296 for sampling pitch.
The primary apparatus is a Mettler Toledo dropping point cell, which consists of a control unit with digital temperature indicator, an integrated or separate furnace, sample cartridges, and accessories. The control unit automatically regulates the heating rate of the furnace. The softening point is indicated on the readout when the sample flow triggers the softening point detection system. General views of the instrument are shown in Fig. 1 (old instrument) and Fig. 2 (new instrument).
The control unit provides a continuous, linear temperature increase from 25°C to 250°C at a rate of 2°C/min, with a digital readout indicating temperature to 0.1°C. The furnace unit can heat one or two sample cup assemblies at a linear rate of 2°C/min ± 0.3°C/min and includes a sensing system capable of detecting the softening point with a precision of 0.1°C.
In the test method, the softening point is defined as the temperature at which a pitch sample, suspended in a cylindrical cup with a 6.35 mm hole at the bottom, flows downward a distance of 19 mm to interrupt a light beam, as the sample is heated at a linear rate in air.
The test method requires precise temperature control and detection. The following table summarizes the essential specifications from the standard.
| 🛠️ Aspect | 📏 Specification |
|---|---|
| Softening Point Range | 50°C to 180°C |
| Heating Rate | 2°C/min ± 0.3°C/min |
| Temperature Indication Precision | 0.1°C |
| Detection Precision | 0.1°C |
| Sample Cup Hole Diameter | 6.35 mm |
| Flow Distance | 19 mm |
These specifications ensure that the softening point determination is consistent and reproducible, with results comparable to other methods in specified temperature ranges.
The softening point is defined as the temperature at which pitch flows 19 mm to interrupt a light beam when heated at 2°C/min.
A linear rate of 2°C/min ± 0.3°C/min ensures consistent viscosity changes and reproducible softening point values.
The digital readout and sensing system provide a precision of 0.1°C throughout the test.
ASTM A314, D2319, and D4296 are referenced for sampling and alternative testing.