Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
This test method, designated ASTM D2866-11, determines the total ash content of activated carbon. The sample must be dried to constant weight (change ≤0.5 mg) at 150 ± 5°C, with 3 hours typically sufficient. For carbons that ignite spontaneously at this temperature, moist carbon should be used with a moisture correction applied per ASTM D2867. All crucibles are ignited at 650 ± 25°C for 1 hour in a muffle furnace, then cooled in a desiccator before use.
| 🛠️ Equipment | ⚡ Specification |
|---|---|
| Muffle Furnace | 650 ± 25°C, air circulation |
| High-Form Crucible | High-temperature resistant |
| Analytical Balance | Sensitivity: 0.1 mg |
| Desiccator | For cooling to ambient temperature |
| Forced-Air Oven | 145–155°C |
Weigh an aliquot of the dried sample into the tared crucible such that the expected ash residue is around 0.1 g. Place the crucible in the muffle furnace at 650 ± 25°C with air circulation. Ash the sample for 3 to 16 hours, until constant weight (within ±0.5 mg) is achieved. After ashing, cool the crucible in a desiccator to ambient temperature and weigh to the nearest 0.1 mg. Automated methods like thermogravimetric analysis (TGA) may be used as alternatives, with furnace temperature set at 650 ± 25°C per ASTM D7582.
The total ash content (%) is calculated as (weight of residue / weight of dried sample) × 100. Report the result to one decimal place. The muffle furnace method is the reference test method, and alternative procedures must be validated against it.
| 📏 Parameter | 🎯 Specification |
|---|---|
| Furnace Temperature | 650 ± 25°C |
| Drying Temperature | 150 ± 5°C |
| Constant Weight Criterion | ±0.5 mg |
| Balance Sensitivity | 0.1 mg |
| Sample Ash Goal | ~0.1 g |
It measures the inorganic residue after complete combustion, which affects the carbon’s adsorptive and catalytic properties.
Drying removes moisture that could bias the weight measurement, with 3 hours typically achieving constant weight.
Use moist carbon and apply a moisture correction using ASTM D2867, then start ashing from a cold furnace.
The muffle furnace procedure at 650 ± 25°C is the reference; TGA methods described in ASTM D7582 are valid alternatives.