D2772-90 – Standard Test Method Technical Guide

🧪 Apparatus Specifications for MgO Sieve Analysis

This test method, designated ASTM D 2772 – 90 (Reapproved 1997), defines the standard procedure for determining the particle size distribution in a representative sample of granular electrical grade magnesium oxide. The required apparatus includes a mechanical sieve shaker, a precise timer, and a triple-beam balance with 0.1 g resolution. A nonmetallic brush and a small rubber rod or mallet are specified for gentle cleaning and sieving assistance.

🚩 Sieve Category 📐 Mesh Sizes (ASTM E 11, Half-Height)
✅ Required Sieves 40, 60, 80, 100, 140, 200, 325 (plus pan and cover)
🔲 Optional Sieves 35, 45, 50, 120, 170

A robust calibration program is essential. Each testing station must maintain two sets of sieves: one active set for routine analysis and a dedicated calibration set reserved exclusively for performance verification.

⚙️ Test Procedure and Calibration Regimen

The sieves are nested with the coarsest mesh on top and the collection pan on the bottom. A representative sample between 100 and 125 g is obtained following Test Method D 2755 and weighed to the nearest 0.1 g. After the mechanical sieving cycle, the mass retained on each sieve is collected.

⚠️ Cleaning Precautions: The standard explicitly prohibits the use of wire brushes for sieve maintenance. Approved cleaning methods include a trichloroethylene degreaser, ultrasonic cleaning, or gentle tapping with a small rubber mallet (Section 5.6).
⚙️ Parameter 📐 Specification / Limit
Sample Mass 100 – 125 g
Balance Resolution 0.1 g
Calibration Frequency Every 50 sieve analyses
Calibration Tolerance ±3 % by weight (vs. calibration set)
Mandatory Sieve Check Every 200 uses or 6 months (whichever first)
💡 Calibration Best Practice: Sieve openings must be verified against a new sieve using a 100-g comparison sample. The active set must meet a tolerance of ±3% by weight compared to the calibration standard. Sieves must also be rechecked immediately if they become slack or clogged (Section 5.4).

📊 Particle Size Distribution and Reporting

After the sieving operation, the mass of MgO retained on each sieve and the bottom pan is weighed to the nearest 0.1 g. The retained mass on each sieve is expressed as a percentage of the total initial sample mass. This cumulative distribution data provides the complete granular profile required for quality control in electrical grade magnesium oxide insulation materials. Per Section 1.2, all values must be reported in acceptable metric units.

❓ Frequently Asked Questions

🔍 What sieves are strictly required for this test method?

According to Section 4.2.1, the required sieves are U.S. Standard mesh sizes 40, 60, 80, 100, 140, 200, and 325, along with a bottom pan and a

📥 Standard Documents Download

🔒
Please wait 10 seconds, the download links will appear after the ad loads

Leave a Reply

Your email address will not be published. Required fields are marked *