Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
ASTM D1657-22, officially designated as the Manual of Petroleum Measurement Standards (MPMS), Chapter 9.2, defines the standard test method for determining the density or relative density of light hydrocarbons, including liquefied petroleum gases (LPG). The method is specifically applicable to materials with Reid vapor pressures exceeding 101.325 kPa (14.696 psi).
A critical safety and technical constraint is the vapor pressure limit of 1.4 MPa (200 psi) at the test temperature for the prescribed apparatus. The standard strictly warns against using the equipment beyond this pressure limit, as it is dictated by the specific type of equipment. SI units are regarded as the standard for all values, with US Customary units provided in parentheses for informational purposes only. The user bears full responsibility for establishing safety and regulatory compliance.
The initial readings obtained from the pressure hydrometer are termed uncorrected hydrometer readings and are not final density measurements. The user must apply a rigorous multi-step correction procedure outlined in Section 11 of the standard. This includes corrections for the meniscus effect, the thermal glass expansion effect, and any alternate calibration temperature effects.
Correction to the reference temperature is performed by means of calculations and standards such as the Adjunct to D1250 Guide (API MPMS Chapter 11.1) or API MPMS Chapter 11.2.4 (GPA TP-27). These fully corrected values are then suitable for use in subsequent calculations, including meter factor correction and base prover volume determination. For converting density to relative density or alternative units, the standard references the Interconversion Procedures in API MPMS Chapter 11.5. Equipment verification and certification must follow the detailed procedure in Annex A1.
Proper application of this standard requires proficiency in several companion industry documents. The tables below summarize the critical parameters and referencing standards required for compliance.
| 🔍 Parameter | 📏 Value / Specification | 📄 Notes / Reference |
|---|---|---|
| Standard Designation | D1657 − 22´1 | Editorially corrected in March 2023 |
| Max Vapor Pressure (Apparatus) | 1.4 MPa (200 psi) | Strict limit based on equipment design |
| Min Reid Vapor Pressure (Sample) | > 101.325 kPa (14.696 psi) | Defines applicability to light hydrocarbons / LPG |
| Primary Unit System | SI Units (Standard) | US Customary values are for information only |
| Required Corrections | Meniscus, Thermal Glass, Calibration Temp | Detailed in Section 11 calculations |
| Equipment Certification | Annex A1 | Formal procedure for verification/certification |
| 📋 Document ID | 📎 Full Title / Description | ⚙️ Application in D1657 |
|---|---|---|
| D1250 / API MPMS Ch 11.1 | Temperature and Pressure Volume Correction Factors | Reference temperature corrections |
| D1265 | Practice for Sampling LP Gases, Manual Method | Recommended sampling practice |
| D1298 / API MPMS Ch 9.1 | Hydrometer Method for Density, Relative Density, and API Gravity | Related hydrometer methodology |
| API MPMS Ch 11.2.4 | Density/Relative Density Conversion (GPA TP-27) | Alternate calculation for corrections |
| API MPMS Ch 11.5 | Interconversion Procedures | Converting between units or reference temperatures |
🔍 What is the official scope of ASTM D1657-22?
It covers the determination of density or relative density of light hydrocarbons, including liquefied petroleum gases (LPG), that have a Reid vapor pressure exceeding 101.325 kPa (14.696 psi).
⚡ What is the maximum vapor pressure allowed for the apparatus?
The prescribed apparatus is limited to materials with vapor pressures lower than 1.4 MPa (200 psi) at the test temperature, as dictated by the specific equipment design.
📌 What corrections must be applied to the initial hydrometer reading?
Corrections must be applied for the meniscus effect, thermal glass expansion, alternate calibration temperature effects, and the reading must be converted to the reference temperature using API MPMS Chapter 11.1 or 11.2.4.
💡 How are values from this test method converted to other units?
Values determined as density or relative density can be converted using API MPMS Chapter 11.5 (Interconversion Procedures), or the Adjunct to D1250 Guide / API MPMS Chapter 11.2.4, as applicable.