Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
Physical Address
304 North Cardinal St.
Dorchester Center, MA 02124
CAN CSA C22.2 No. 60601-2-33-12 (2016) is the Canadian adoption of the international standard IEC 60601-2-33:2010, with amendments specific to the Canadian regulatory framework. This standard applies to the basic safety and essential performance of magnetic resonance equipment (MRE) intended for medical diagnosis. It covers all active devices used for magnetic resonance imaging (MRI), including systems, coils, and associated accessories that are necessary for clinical operation.
The standard addresses hazards unique to MRI technology, such as static magnetic fields, time-varying gradient magnetic fields, radio-frequency (RF) electromagnetic fields, cryogenic liquids, and acoustic noise. It does not cover systems intended for animal imaging or laboratory use, nor does it apply to magnetic resonance spectroscopy systems unless they are used for diagnostic imaging.
The standard imposes strict limits on static magnetic field exposure for patients, operators, and the general public. For the public, the static magnetic field must not exceed 0.5 mT at any location accessible to untrained persons. For trained operators, the limit is 2 T unless additional precautions are implemented. The fringe field must be controlled through magnetic shielding and access restriction.
Systems using superconducting magnets must incorporate a quench handling mechanism that safely vents cryogenic gas (typically helium) to the outdoors. Pressure relief devices and rupture disks must comply with CSA B51 requirements. The quench vent must be sized to handle the maximum boil-off rate, and the enclosure must prevent accumulation of asphyxiant gases.
Radio-frequency energy is limited to avoid excessive heating of patient tissue. The standard specifies whole-body SAR (specific absorption rate) limits of 4 W/kg averaged over 15 minutes for normal mode operation, 2 W/kg for first-level controlled mode, and 1 W/kg for second-level controlled mode. Gradient field characteristics must limit peripheral nerve stimulation; the dB/dt (time derivative of the magnetic flux density) must remain below the established thresholds for both normal and controlled operating modes.
| Parameter | Requirement | Test Condition |
|---|---|---|
| Static field (public access) | ≤ 0.5 mT | At any accessible point |
| Whole-body SAR (Normal Mode) | ≤ 2 W/kg (15 min average) | Head or body coil |
| Gradient slew rate (Normal Mode) | ≤ 20 T/m/s (as defined in IEC 60601-2-33) | For typical coil |
| Acoustic noise (peak) | ≤ 99 dBA (Normal Mode) | At operator position |
| Quench vent flow | Able to handle maximum helium boil-off | Simulated quench test |
The standard defines essential performance as the ability to produce diagnostic-quality images while maintaining patient safety. Essential performance includes SNR (signal-to-noise ratio) stability, spatial accuracy, and slice positioning accuracy. The manufacturer must declare tolerances and demonstrate that no single fault condition results in a complete loss of essential performance.
To achieve compliance, manufacturers must perform a comprehensive risk management process in accordance with ISO 14971. The manufacturer must identify all hazards related to the MRI environment, including ferromagnetic projectile accidents, cryogen spills, and gradient-induced nerve stimulation. Each hazard must be evaluated and mitigated to an acceptable residual risk.
Testing is typically performed on a representative system configuration. Required tests include:
Each unit must bear a nameplate with the manufacturer name, model, serial number, date of manufacture, and the CSA mark. Operator manuals must include warnings about implanted devices, magnetic field hazards, and emergency procedures. Additionally, a technical description must specify static field strengths, gradient characteristics, and quench vent requirements for installation.
While the core requirements align with IEC 60601-2-33:2010, the 2016 CSA edition introduced clarifications on:
These modifications ensure that MRI equipment installed in Canada meets the same high level of safety as other medical electrical devices while accommodating local infrastructure and regulatory expectations.
Article published for informational purposes. Always consult the latest published edition of the standard and a qualified certification body for official compliance guidance. © 2026