🩺 BS EN ISO 22442 is the British Standard series on medical devices utilizing animal tissues and their derivatives, classified under ICS 11.100.99 (Biological evaluation of medical devices) and 11.100.20 (Sterilization of medical devices). Medical devices made from animal-derived tissues (such as bovine pericardium, porcine small intestinal submucosa, and bovine/porcine bone) and their derivatives (such as collagen and gelatin) are widely used in cardiovascular, orthopaedic, wound care, and surgical repair applications. Such devices carry specific risks due to their animal-derived material, including viral and prion transmission and immunogenicity. The series specifies the application of risk management (Part 1) and controls on sourcing, collection, and handling (Part 2), providing a systematic approach to safety assessment and production control of animal-derived medical devices.
💡 The risks of animal-derived medical devices differ from those of conventional devices: beyond physical and chemical risks, they involve pathogen transmission (viruses, prions), immunogenic reactions, and traceability of animal origin. Risk management and sourcing controls are the two pillars ensuring safety.
1. Scope and General Framework 📌
The series was prepared by ISO/TC 194 (Biological evaluation of medical devices) in collaboration with CEN/TC 316 (Medical devices utilizing tissues), with UK participation through Technical Committee RGM/1 (Regenerative medicine). The series is structured as follows:
| Standard Part | Year | Subject | Main Content |
| BS EN ISO 22442-1 | 2015 | Application of risk management | Risk management requirements and methods for animal-derived medical devices |
| BS EN ISO 22442-2 | 2015 | Controls on sourcing, collection, and handling | Control requirements for animal sourcing, tissue collection, and handling |
⚠️ Animal-derived medical devices are implanted in or contact the human body; the origin and processing of raw materials are critical to safety. Control the entire chain — from animal species, region, health status, tissue collection and transport, to virus inactivation/removal processes — and maintain complete traceability records.
2. Series Structure and Technical Points 🔬
2.1 Part 1 Application of Risk Management (BS EN ISO 22442-1:2015)
Part 1 specifies the application of risk management for medical devices manufactured from animal tissues and their derivatives, superseding the withdrawn EN ISO 22442-1:2007. Building on ISO 14971 (Application of risk management to medical devices), it identifies, analyses, and controls risks specific to animal-derived materials (viral and prion transmission, immunogenicity, bioburden, toxic residues, material performance, and biocompatibility). It requires manufacturers to establish a risk management process and, combined with virus inactivation/removal validation, animal origin controls, and production process controls, to reduce risks to an acceptable level.
2.2 Part 2 Controls on Sourcing, Collection and Handling (BS EN ISO 22442-2:2015)
Part 2 specifies the control requirements for sourcing, collection, and handling of animal tissues and their derivatives used to manufacture medical devices, superseding the withdrawn EN ISO 22442-2:2007. It covers: determination of animal origin and species, animal health and quarantine requirements, tissue collection and transport conditions, tissue processing requirements (such as disinfection, inactivation, and virus removal processes), and traceability (from animal to finished product) and record management. Through rigorous process control, the safety and consistency of animal-derived materials are ensured and pathogen transmission prevented.
💡 The series has a clear relationship with EU medical device directives/regulations: EN ISO 22442 provides an annex mapping to the Essential Requirements of EU Directive 93/42/EEC, and is an important harmonised standard basis for safety assessment of animal-derived medical devices in EU CE certification.
3. Key Technical Requirements 🔧
The series sets out the following key technical requirements for animal-derived medical devices:
- Risk management process: establish a risk management system per ISO 14971, identifying animal-derived-specific risks and applying control measures.
- Animal origin controls: define animal species, source region, and health status, with quarantine and health certification where necessary.
- Virus inactivation/removal: validate virus inactivation and removal processes for animal-derived materials to ensure viral safety.
- Traceability: establish a complete traceability system from animal origin and tissue batch to finished product batch, supporting problem tracing and recall.
- Biocompatibility and immunogenicity: evaluate material biocompatibility per relevant standards and control immunogenic risk.
🚨 The viral and prion risks of animal-derived materials have public health significance, particularly the risk of transmissible spongiform encephalopathies (TSE) such as bovine spongiform encephalopathy (BSE). Manufacturers should strictly control animal origin and risk grading, and follow relevant regulatory requirements.
4. Engineering Practice and Application Guidance 🛠️
- 🔍 Product development: conduct animal-derived material risk assessment early in development, defining risk control strategies and validation plans.
- 🔧 Production control: establish standard operating procedures (SOPs) for animal sourcing, tissue collection, and processing, strictly executing virus inactivation and removal processes.
- 📊 Supply chain management: qualify and health-monitor animal origin suppliers to ensure material traceability.
- 📋 Regulatory compliance: for EU and other markets, complete safety assessment and CE certification documentation per harmonised standards and regulatory requirements.
5. Frequently Asked Questions ❓
❓ Q: How do the risks of animal-derived medical devices differ from conventional devices?
A: Beyond physical and chemical risks, they involve specific risks such as viral and prion transmission, immunogenic reactions, bioburden, and toxic residues, hence the need for dedicated risk management and sourcing controls.
❓ Q: Why is animal origin control so important?
A: Animal health status and source region directly affect the risk of carrying pathogens (such as viruses and prions). Strict origin controls and quarantine significantly reduce the possibility of pathogens entering the medical device.
❓ Q: What is virus inactivation/removal validation?
A: It refers to demonstrating through dedicated tests that inactivation/removal processes effectively reduce or eliminate viruses in the material, a key step in confirming viral safety.
❓ Q: What is the relationship between this series and EU regulations?
A: EN ISO 22442 provides an annex mapping to the Essential Requirements of EU medical device directives (such as 93/42/EEC), and is an important basis for CE certification and safety assessment of animal-derived medical devices.