📡 BS EN ISO 18563 is the British Standard series on characterization and verification of ultrasonic phased array (Phased Array Ultrasonic Testing, PAUT) equipment, classified under ICS 19.100 (Non-destructive testing). As phased array ultrasonics finds growing application in inspecting industrial welds, pipelines, forgings, and composites, objectively and comparably characterizing and verifying the performance of phased array instruments and combined systems has become a key industry concern. The series provides a unified basis for procurement acceptance, periodic verification, and capability confirmation of phased array equipment, supporting reliable inspection results and mutual recognition of test data.
💡 Phased array ultrasonics electronically controls the excitation timing of individual elements in an array transducer to focus and steer the beam, covering a large area and producing multi-angle imaging in a single scan. Its performance verification is far more complex than that of a conventional single-probe ultrasonic test, hence the need for dedicated characterization standards.
1. Scope and General Framework 📌
The series was prepared by ISO/TC 135 (Non-destructive testing) in collaboration with CEN/TC 138 (Non-destructive testing), with UK participation through Technical Committee WEE/46 (Non-destructive testing). The series specifies characterization and verification requirements at two levels — the individual instrument and the complete system:
| Standard Part | Year | Subject | Object of Verification |
| BS EN ISO 18563-1 | 2015 | Instruments | Phased array ultrasonic instrument (main unit / electronics) |
| BS EN ISO 18563-3 | 2015 | Combined systems | Complete inspection system of instrument + phased array probe + connecting cable |
⚠️ The performance of a phased array system depends on the combined effect of the instrument electronics, the array probe, and the connecting cable. Verifying the instrument alone does not guarantee the capability of the whole system, so combined system verification in Part 3 is indispensable.
2. Series Structure and Technical Points 🔬
2.1 Part 1 Instruments (BS EN ISO 18563-1:2015)
Part 1 specifies methods for characterizing and verifying phased array ultrasonic instruments, applicable to multi-channel phased array instruments. Verification typically covers: pulse-generator/receiver performance consistency across channels (elements), inter-channel consistency, time-gain compensation (TGC) accuracy, pulse width and amplitude, receiving bandwidth, noise level, and the instrument response to different delay and focusing modes. Using standardised test loads and measurement procedures, the electrical and acoustic performance of the instrument can be quantitatively assessed, ensuring comparable performance between instruments.
2.2 Part 3 Combined Systems (BS EN ISO 18563-3:2015)
Part 3 specifies methods for characterizing and verifying the complete inspection system comprising a phased array instrument, an array probe, and a connecting cable. It accounts for the combined effect of the transducer’s acoustic field, the instrument’s electronic performance, and the overall system response, using standard reference blocks (such as comparison blocks with flat-bottom holes or side-drilled holes) to verify focusing, sensitivity, resolution, and imaging quality. Combined system verification addresses the performance of the whole equipment as used in practice, reflecting field capability better than verification of the instrument alone.
💡 In practice one typically “verifies the instrument first, then the system”: confirm the instrument’s electronic performance per Part 1, then verify the whole system’s detection capability on standard reference blocks per Part 3, forming a complete performance confirmation chain.
3. Key Technical Requirements 🔧
The series sets out the following key technical requirements for verifying phased array equipment:
- Channel consistency: the transmit and receive performance of all elements should be as uniform as possible, with inter-channel differences within the permitted range.
- Standard reference blocks: use reference blocks with known geometry and reflectors (e.g., flat-bottom holes, side-drilled holes, or slots) as the verification basis to ensure traceable results.
- Beam characteristics: verify acoustic parameters such as focal depth, steering angle, and beam width to confirm they meet the inspection procedure requirements.
- Sensitivity and resolution: the system should have sufficient detection sensitivity and defect resolution capability to meet the acceptance criteria of the inspected object.
- Repeatability: repeated measurements under the same conditions should remain stable, and the system should have good long-term stability.
🚨 The reliability of phased array inspection results depends on verifiable equipment performance. If equipment is unverified or fails verification, missed detection or misjudgement may result, posing a serious threat to structural safety. Always complete performance confirmation and keep verification records before use.
4. Engineering Practice and Application Guidance 🛠️
- 🔍 Equipment procurement acceptance: on delivery, verify phased array equipment performance per Part 1 and Part 3, forming procurement technical conditions and acceptance records.
- 🔧 Periodic verification: establish a periodic equipment performance verification regime; re-verify system performance before testing, during testing, or after major repair.
- 📊 Inspection procedure development: based on verified system performance, set inspection sensitivity, scan range, and acceptance criteria so that procedures match actual equipment.
- 📋 Personnel training: operators should understand phased array principles and equipment verification methods, and be able to independently perform performance confirmation and result interpretation.
5. Frequently Asked Questions ❓
❓ Q: How does BS EN ISO 18563 differ from conventional ultrasonic test equipment standards?
A: Conventional ultrasonic test equipment is typically single-channel and relatively simple to verify; phased array equipment is multi-channel and requires verification of phased-array-specific performance such as channel consistency, focusing, and steering, hence the need for dedicated standards.
❓ Q: Why is combined system verification in Part 3 important?
A: Because the combination of instrument, probe, and cable introduces additional performance variation; a qualified instrument alone does not guarantee a qualified system. Combined system verification ensures the actual inspection equipment meets requirements.
❓ Q: What role do standard reference blocks play in verification?
A: Reference blocks provide known reflectors and a reference basis for measuring sensitivity, resolution, focusing, and imaging quality, and are the foundation for traceable and comparable verification results.
❓ Q: How should the verification interval be determined?
A: The interval should consider equipment usage frequency, environmental effects, and inspection task requirements; typically periodic checks (e.g., annually or semi-annually) are recommended, with immediate re-verification after repair, component replacement, or anomalies.