🧪 BS EN ISO 15587 is a British Standard series on sample digestion in water analysis, prepared by ISO/TC 147 (Water quality) with CEN/TC 230 (Water analysis) and published in the UK within the BS 6068 water-quality series. It defines methods for the digestion pre-treatment of water samples so that selected elements such as cadmium, chromium, copper, lead, nickel and zinc can be determined by atomic absorption spectrometry (AAS), inductively coupled plasma optical emission spectrometry (ICP-OES) or inductively coupled plasma mass spectrometry (ICP-MS). The series has two parts: Part 1 specifies the aqua regia digestion method and Part 2 specifies the nitric acid digestion method.
💡 The purpose of digestion is to convert particulate, colloidal or complexed metals in the sample into a dissolved form measurable by the instrument, while destroying the organic matrix. Which digestion method is chosen depends on the elements concerned, the sample matrix and the analytical objective.
1. Overview of the Series 📌
Classified under ICS 13.060.50 (Water quality — examination of chemical properties of water) and published in 2002, the series is part of the BS 6068 water quality standard family. UK participation is entrusted to Technical Committee EH/3 (Water quality) through Subcommittee EH/3/2 (Physical, chemical and biochemical methods). Both parts carry “Incorporating Corrigendum No. 1” and include Annex ZA (relationship with the relevant EU directive).
| Standard Number | Title | Digestion Reagent | Typical Use |
BS EN ISO 15587-1:2002 (BS 6068-2.76:2002) | Water quality — Digestion for the determination of selected elements in water — Part 1: Aqua regia digestion | Hydrochloric + nitric acid (aqua regia) | Total metal content, various matrices |
BS EN ISO 15587-2:2002 (BS 6068-2.77:2002) | Water quality — Digestion for the determination of selected elements in water — Part 2: Nitric acid digestion | Nitric acid | Dissolved / more easily digested metals |
⚠️ Digestion involves strong acids and high temperatures. Operators must have appropriate chemical analysis experience and strictly comply with laboratory safety and waste-disposal requirements. The Standard itself stresses that these tests should be performed only by suitably qualified personnel with an appropriate level of chemical expertise.
2. Structure and Technical Essentials 🔧
The principal difference between the two parts is the digestion reagent and the resulting digestion strength: aqua regia digestion (Part 1) uses a mixture of concentrated hydrochloric and nitric acids, which is more powerfully oxidizing and complexing, dissolving most metals (including some hard-to-dissolve particulate metals); nitric acid digestion (Part 2) uses nitric acid alone, which is less oxidizing and suits samples that are easier to digest or where only dissolved metals are of interest. Both methods require heating under reflux so that volatile elements are not lost.
| Comparison | Part 1 — Aqua Regia | Part 2 — Nitric Acid |
| Reagent | Hydrochloric + nitric acid (aqua regia) | Nitric acid |
| Digestion strength | Strong; dissolves most metals | Relatively weaker |
| Matrices suited | Wastewater, sludge, complex samples | Clean water, simpler matrices |
| Reflux requirement | Heat under reflux to prevent loss | Heat under reflux to prevent loss |
| Determination | AAS / ICP-OES / ICP-MS | AAS / ICP-OES / ICP-MS |
3. Key Technical Requirements: Samples and Reagents 🔬
The Standard places clear requirements on sample collection, preservation and reagent purity to ensure reliable results:
- Sample preservation: samples should be analysed as soon as possible after collection, or acidified and stored under defined conditions; container material should avoid adsorption and contamination — polyethylene or PTFE containers are commonly used.
- Reagent purity: the acids used should be analytical reagent (AR) grade or higher to avoid introducing the test metals and raising blank values; water should be high-purity (deionised/ultra-pure) water.
- Blank control: a reagent blank should be prepared with each digestion batch to correct for contamination from reagents and vessels.
- Vessel cleaning: digestion vessels should be acid-washed and rinsed with high-purity water to avoid cross-contamination and residues.
✅ The Standard recommends inserting certified reference materials (CRM) or spike-recovery tests in each batch to verify the accuracy of digestion and determination; recoveries should fall within the specified range, otherwise the cause must be investigated and the analysis repeated.
4. Key Technical Requirements: Digestion Procedure 🎯
The basic procedure for both digestion types is similar — “sample → add reagent → heat under reflux → cool → make up to volume”:
- Sample volume: take an appropriate sample volume based on element concentration and instrument sensitivity, into a reflux flask.
- Acid addition: aqua regia digestion adds hydrochloric and nitric acid in proportion; nitric acid digestion adds nitric acid only. Acid addition should be performed in a fume hood.
- Reflux heating: heat slowly to boiling on a reflux apparatus and hold for the specified time until complete digestion; the reflux condenser prevents loss of acid fumes and volatile elements.
- Cooling and make-up: after cooling, the digest is transferred and made up to the specified volume, filtered if necessary to remove insolubles, and then subjected to instrumental determination.
🚨 Digestion must comply strictly with laboratory safety rules: strong acids are highly corrosive, and heating may generate acid fumes and hazardous gases. Always work in a fume hood and wear goggles, acid-resistant gloves and protective clothing.
5. Engineering Practice and Applications 🛠️
- 🔍 Method selection: in environmental monitoring, drinking water and surface water analysis, aqua regia digestion is normally used when “total metals” are required; when only dissolved metals are of interest, nitric acid digestion or direct filtration before analysis may be used.
- 📊 Instrument matching: digested samples are quantified by AAS (flame or graphite furnace), ICP-OES or ICP-MS according to the element and concentration range.
- 📋 Quality control: establish a complete QC programme including blanks, duplicate samples, spike recoveries, certified reference materials and control charts to ensure traceable and reproducible data.
- 🧪 Typical applications: the series is widely used for monitoring drinking water, surface water, groundwater, domestic sewage and industrial effluent, and for metal-element determination in environmental assessment and pollution-source investigations.
6. Frequently Asked Questions ❓
❓ Q: Can the results of aqua regia and nitric acid digestion be used interchangeably?
A: Not necessarily. The two methods differ in digestion strength and in their ability to dissolve particulate and less-soluble metals. Aqua regia digestion is usually used for “total metals”; nitric acid results are comparable only when the matrix is confirmed to be simple and the elements easily digested.
❓ Q: Why must digests be made up to volume and filtered?
A: Making up to volume provides a defined volume basis for concentration calculation; filtration removes undissolved particles and suspended matter, avoiding nebulizer blockage and instrumental interference.
❓ Q: Why heat under reflux instead of an open vessel?
A: Reflux reduces losses of acid fumes and volatile elements (e.g., mercury, arsenic), allowing full digestion under closed reflux and ensuring recovery and accuracy. Open heating can lose volatile elements and give low results.
🔍 Q: How do you judge whether digestion is complete?
A: Completeness is judged mainly by the digest becoming clear with no visible particles and stable colour, combined with spike-recovery rates and certified reference material results; if digestion is incomplete, adjust the acid amount, heating time or switch to a stronger digestion mode.