🔥 BS EN ISO 11990 is the BSI implementation of the EN ISO 11990 series, which defines test methods for determining the ability of a critical airway-management device — the tracheal tube — to resist laser irradiation. During laser surgery of the larynx, trachea, or bronchi (e.g., laryngeal tumour resection, vocal-cord procedures), the laser beam may inadvertently strike an already-inserted tracheal tube. If the tube material is not laser-resistant, it can be punctured or ignited, causing airway fire, tissue burns, and other serious complications. The series has two parts: Part 1 addresses the tube shaft, and Part 2 the tube cuff. It was prepared by ISO/TC 172 (Optics and photonics) with CEN/TC 123 (Lasers and photonics), classified under ICS 11.040.10 (Surgical instruments and materials) and 31.260 (Laser equipment).
💡 Laser-induced airway fire is one of the most dangerous complications in anaesthesia and ENT surgery. BS EN ISO 11990 provides a unified test method that helps manufacturers develop “laser-resistant” tubes and helps clinical teams evaluate and select products.
1. Scope and General Framework 📌
The “laser resistance” covered by this series is the ability of a tracheal tube to resist ignition and penetration under specific laser parameters (wavelength, power density, exposure time, beam angle, and beam diameter). The standard defines systematic test methods used to:
- uniformly evaluate the laser resistance of existing and new tube materials and constructions;
- provide manufacturers with a basis for product development and quality control;
- give clinicians an objective comparison and selection reference.
The scope of the standard includes:
- Shaft (Part 1): penetration and ignition behaviour of the tube wall and shaft material under laser irradiation.
- Cuff (Part 2): rupture, leakage, and ignition behaviour of an inflated cuff under laser irradiation.
- Laser types: the clinically common continuous-wave and pulsed lasers, such as CO₂, Nd:YAG, KTP, and diode lasers.
⚠️ No tracheal tube is absolutely “laser-resistant” under arbitrary laser parameters. The standard evaluates resistance under controlled, specified test conditions; clinical use still requires safe operating practices such as the lowest practicable inspired oxygen concentration and minimum power.
2. Series Composition and Sub-parts 🗂️
| Standard No. | Title | Published | Main Technical Content |
| BS EN ISO 11990-1:2014 | Lasers and laser-related equipment — Determination of laser resistance of tracheal tubes — Part 1: Tracheal tube shaft | October 2014 | Specifies test apparatus, irradiation method, and criteria for shaft laser resistance |
| BS EN ISO 11990-2:2014 | Lasers and laser-related equipment — Determination of laser resistance of tracheal tubes — Part 2: Tracheal tube cuffs | October 2014 | Specifies test apparatus, irradiation method, and criteria for inflated-cuff laser resistance |
The two parts share a common structure and are based on ISO 11990-1:2011 and ISO 11990-2:2010; the British Standards were published by BSI on 30 November 2014. Earlier versions (BS EN ISO 11990-1:2011 and BS EN ISO 11990-2:2010) were superseded and withdrawn.
3. Part 1: Determination of Shaft Laser Resistance 🔬
The shaft test simulates the accidental exposure of the tube wall to a laser beam. Key technical points include:
- Test apparatus: the shaft is secured in a defined rig, and the beam is directed at the wall at a specified angle (usually normal or slightly oblique incidence).
- Laser parameters: the test is carried out at an agreed wavelength, output power, spot diameter, and exposure time, distinguishing continuous-wave and pulsed operation where necessary.
- Gas flow conditions: air or an oxygen-containing gas is passed through the tube at the specified rate, reproducing the clinical breathing-circuit environment.
- Criteria: the time to penetration (from start of irradiation to breakthrough) and any ignition are recorded; laser resistance is characterised by whether the tube fails to penetrate and ignite within the specified time.
- Test report: tube specification, laser parameters, and penetration/ignition results are recorded for comparison and certification.
🚨 If penetration or ignition occurs during shaft testing, stop the irradiation immediately, cut off the gas flow, and initiate the extinguishing procedure. Testing must be carried out in a laboratory with laser safety protection and fire-fighting provisions.
4. Part 2: Determination of Cuff Laser Resistance 🎈
The cuff is the part of the tube in direct contact with airway tissues and is typically close to the surgical site — a weak point. Part 2 covers:
- Inflation state: the cuff is inflated to the manufacturer’s specified pressure or volume.
- Irradiation mode: the laser is directed at the surface of the inflated cuff with defined parameters.
- Principal risks: laser puncture of the cuff causes gas leakage and loss of seal on one hand, and the escaping gas may ignite on the other.
- Criteria: the time to puncture/leakage is recorded, evaluating the resistance of cuff material and construction.
- Safety significance: laser-resistant cuffs (e.g., flame-retardant materials or metal-foil reinforcement) significantly reduce airway-fire risk in laser surgery.
5. Test Conditions and Result Evaluation 🔧
| Test Element | Typical Values / Note | Effect on Result |
| Laser wavelength | CO₂ (10.6 μm), Nd:YAG (1064 nm), KTP (532 nm), etc. | Absorption mechanisms differ by wavelength, directly affecting penetration behaviour |
| Output power / power density | Set within clinically common ranges | Higher power penetrates faster |
| Spot diameter and incidence angle | Defined beam diameter and angle | Affects local energy density |
| Gas flow / oxygen concentration | Air or low-concentration oxygen flow | Determines ignition risk |
| Criteria | No penetration, no ignition within the specified time | Gives “pass/fail” or time-to-penetration |
✅ Standardised test parameters and criteria make results comparable across manufacturers and laboratories; the time to penetration and whether ignition occurred in the test report are important objective inputs for clinical selection.
6. Engineering Practice and Applications 🛠️
- 🏥 Clinical selection: for laser surgery of the trachea, bronchi, and larynx, prefer tubes validated to this standard with clearly demonstrated laser resistance.
- 🛡️ Safe operation: even with a laser-resistant tube, keep the inspired oxygen concentration at the lowest safe level, avoid prolonged high-power exposure, and keep extinguishing equipment and saline at hand.
- 🏭 Manufacturing and QC: manufacturers perform shaft and cuff laser-resistance tests on each batch of tubes as a release-inspection item.
- 📋 Product labelling: laser-resistance performance and applicable laser types should be clearly stated in instructions for use and labelling for clinical verification.
- 📚 Education and training: include airway-fire prevention and test-method knowledge in anaesthesia and ENT training curricula.
7. Frequently Asked Questions ❓
❓ Q: Why are only the shaft and cuff tested?
A: The shaft and cuff are the parts most likely to be struck and pose the highest risk in laser surgery: shaft puncture can cause airway fire, and cuff puncture causes seal failure with possible ignition; separate test methods are therefore defined. Other components carry lower risk.
❓ Q: Is a laser-resistant tube absolutely safe?
A: No. The standard only demonstrates resistance under the specified laser parameters and conditions; exceeding the parameters or improper handling can still cause incidents. Clinical use must combine low oxygen concentration, low power, and other comprehensive protective measures.
❓ Q: Can results at one laser wavelength substitute for another?
A: No. Absorption and damage mechanisms differ by wavelength; a material resistant at one wavelength may be punctured quickly at another. Tests must be performed at the wavelength actually used.
❓ Q: What should be done if ignition occurs during testing?
A: Stop the laser immediately, cut off the gas flow, remove the tube, and rinse with saline while initiating the fire emergency plan; test facilities should carry extinguishing equipment and laser protection measures.