๐ BS EN ISO 11140 is the British Standard series for chemical indicators used in the sterilization of health care products, implemented by BSI within the framework of ISO/TC 198 (Sterilization of health care products) and equivalent to the ISO 11140 series. Chemical indicators give a visible change (colour change, deformation, etc.) reflecting the physical and chemical conditions of the sterilization process, and are an important monitoring tool in steam, dry heat, ethylene oxide, formaldehyde and hydrogen peroxide sterilization. The series sets out the classification of chemical indicators, general requirements, and the Class 2 indicator systems and indicators used for steam-penetration monitoring. Together with biological-indicator standards and sterilizer-performance standards, it forms the complete chain of medical sterilization quality assurance.
The series applies to chemical indicators used in the sterilization of health care products, covering classification, performance requirements, test methods and instructions for use. Part 1 gives the general requirements and classification framework; Part 3 specifies Class 2 indicator systems for the Bowie and Dick-type steam penetration test; Part 4 specifies Class 2 indicators used as an alternative to the Bowie and Dick-type test for detecting steam penetration. Used together, they support daily monitoring and release decisions on the steam-penetration performance of sterilizers and on process adequacy.
From a drafting perspective, Part 4 supersedes and withdraws BS EN 867-4:2001 (the former European indicator standard); Part 3 is identical with ISO 11140-3:2007 (including Corrigendum 1:2007).
| Standard | Part Subject | Status | Core Content |
|---|---|---|---|
| BS EN ISO 11140-1:2014 | Part 1: General requirements | 2014 edition, 2015 corrigendum; supersedes 2009 | Classification, grading system, performance requirements and general test methods |
| BS EN ISO 11140-3:2009 | Part 3: Class 2 indicator systems for the Bowie and Dick-type steam penetration test | 2009 edition; identical with ISO 11140-3:2007 | Requirements for Class 2 indicator systems for the steam penetration (Bowie-Dick) test |
| BS EN ISO 11140-4:2007 | Part 4: Class 2 indicators as an alternative to the Bowie and Dick-type test | 2007 edition; supersedes BS EN 867-4:2001 | Class 2 indicators for detecting steam penetration (alternative to the Bowie-Dick test) |
Part 1 establishes the classification system and general requirements for chemical indicators, defining the combination of process parameters reflected by each class (1 to 6) and its purpose: Class 1 process indicators show whether items have been through a sterilization process; Class 2 indicators are for specific tests; Class 3 are single-variable indicators; Class 4 are multi-variable indicators; Class 5 are integrating indicators; and Class 6 are emulating (cycle-verification) indicators. The standard sets requirements for marking, instructions, storage conditions, colour-change performance and batch consistency, and requires manufacturers to provide instructions for use and performance claims.
Part 3 specifies the requirements for Class 2 indicator systems used in the Bowie and Dick-type steam penetration test. The Bowie-Dick test verifies the ability of a pre-vacuum steam sterilizer to remove air under an empty chamber and allow steam to penetrate uniformly. The indicator system comprises an indicator and a test pack (process challenge device); under the specified temperature and time it should show a uniform colour change, while a non-uniform change indicates residual air or poor steam penetration.
Part 4 specifies Class 2 indicators that may be used as an alternative to the Bowie and Dick-type test for detecting steam penetration. These indicators are similarly sensitive to steam penetration but are provided as more convenient standalone indicator items for routine daily monitoring. The standard sets performance, test methods, marking and instructions for use so that the monitoring effect is equivalent to that of the Bowie-Dick test.