๐ BS EN ISO 10101 is the British Standard series that specifies the determination of water content in natural gas by the Karl Fischer method. Prepared by BSI under the direction of the Petroleum Standards Policy Committee, it is the English language version of EN ISO 10101:1998 and is identical with ISO 10101:1993. The series supersedes and renumbers the corresponding clauses of BS 3156-11.3.1 to 11.3.3:1994 (the water-in-natural-gas sections of fuel-gas analysis) and comprises three parts: Part 1 (Introduction), Part 2 (Titration procedure) and Part 3 (Coulometric procedure). Accurate determination of the water content of natural gas is fundamental to custody transfer metering, pipeline transmission, liquefaction, and the safe and economic operation of downstream facilities, which makes this series a cornerstone analytical reference for the natural gas industry.
๐ก The Karl Fischer method is a classic chemical technique for water determination: in the presence of iodine, sulphur dioxide and an organic base such as pyridine, water takes part in a quantitative chemical reaction, and the water content is quantified by titration or by coulometric charge measurement. For a gas matrix such as natural gas, which may contain sulphur compounds and heavier hydrocarbons, the series gives specific provisions for controlling interferences.
1. Series Overview and Scope ๐
The series applies to the determination of water in natural gas, covering the complete chain from sampling to the reporting of results. It sets out the principle of the method, the chemical reactions involved and the interferences to be expected, reagents and apparatus, sampling requirements, operating procedures, expression of results, and the content of the test report. Information on repeatability and reproducibility is also provided, giving laboratories a common basis for inter-laboratory comparisons and quality control.
It is worth noting that this series is the successor to Section 11.3 of the former BS 3156 (Analysis of fuel gases): in 1998 BSI renumbered BS 3156-11.3.1, 11.3.2 and 11.3.3 respectively as BS EN ISO 10101-1, -2 and -3 through amendments 10051, 10052 and 10053. The earlier numbers are no longer published separately.
โ ๏ธ Before use, verify the currently valid edition of the referenced documents: ISO 10101 has been amended and updated since 1993, so engineering and commercial applications should always follow the latest official BSI publication and confirm the edition (or amendment) cited in the contract.
2. Series Structure and Responsibilities of Each Part ๐ง
| Standard | Part Subject | Former BS Number | Core Content |
| BS EN ISO 10101-1:1998 | Part 1: Introduction | BS 3156-11.3.1:1994 | Scope, principle, reactions and interferences, general sampling |
| BS EN ISO 10101-2:1998 | Part 2: Titration procedure | BS 3156-11.3.2:1994 | Reagents, apparatus, standardization of Karl Fischer reagent, titration operation |
| BS EN ISO 10101-3:1998 | Part 3: Coulometric procedure | BS 3156-11.3.3:1994 | Coulometric cell, gas injection, charge-to-mass conversion of results |
The three parts form one complete methodology: Part 1 provides the overall framework and interference-control strategy; Part 2 suits the moderate water-content range, deriving water from the amount of reagent consumed; Part 3 is intended for precise measurements at lower water contents, deriving water from the coulombs consumed in the electrolytic generation of iodine, and offers higher sensitivity.
3. Key Technical Requirements ๐ฌ
3.1 Principle of the Method
The method is based on the Karl Fischer reaction: iodine (Iโ) reacts with sulphur dioxide (SOโ) in the presence of water, and the amount of iodine consumed is proportional to the water content. In the titration procedure, a standardized Karl Fischer reagent of known titer is titrated to the end point and the water is calculated from the volume consumed. In the coulometric procedure, iodine is generated electrochemically at the anode of the titration cell and the water is calculated from the quantity of electricity consumed, in accordance with Faraday’s law, so no daily standardization of the reagent is required.
3.2 Interferences and their Control
- Matrix interferences: reducing components such as hydrogen sulphide present in natural gas may react with iodine and give a positive bias; the series requires specific control at the sampling and analysis stages, with pre-treatment or an appropriate end-point indication where necessary.
- Water adsorption/desorption: sampling lines, vessels and apparatus surfaces adsorb water; the sampling system should be kept dry with proper purging and preconditioning to avoid systematic errors.
- Reagent and apparatus cleanliness: all glassware and connections must be dry; gas inlet systems should use structures such as three-way valves to control flow, and drying tubes should be fitted in the outlet lines to keep moisture out of the system.
3.3 Sampling Requirements
The series gives clear sampling provisions: samples should be taken with a suitable sampler and method so as to guarantee representativeness, and analysis should follow as soon as practicable to avoid water condensing or being lost during pressure release or temperature change.
4. Engineering and Laboratory Practice ๐ ๏ธ
- ๐ Custody transfer metering: natural gas is traded by energy content, and water content affects the calorific value per unit volume and the compression factor; it is one of the key components in the measurement data chain.
- ๐ง Pipeline and facility safety: water is a key factor in hydrate formation and corrosion; long-distance pipelines, compressors and liquefaction plants all need moisture monitoring to prevent hydrate blockage and internal corrosion.
- ๐ Laboratory quality management: repeatability and reproducibility data from the series can be used in proficiency testing and uncertainty evaluation; the test report should include sample information, the part used, the units of the result and the test conditions.
- ๐ Compliance and traceability: BS EN ISO 10101 is used together with related international documents such as ISO 6712 and ISO 7504; procurement, acceptance and dispute resolution should cite the method edition.
โ
Part selection: it is advisable for a general gas-analysis laboratory to maintain both routes โ titration covers the routine content range with straightforward operation, while coulometry serves precise low-moisture measurement; the two cross-check each other and markedly improve confidence in the data.
5. Frequently Asked Questions โ
โ Q: What is the relationship between BS EN ISO 10101 and BS 3156-11.3?
A: BS EN ISO 10101-1/-2/-3:1998 are simply the renumbered British Standards for the former BS 3156-11.3.1 to 11.3.3:1994; they are identical with ISO 10101:1993, and the original numbers are no longer published separately.
โ Q: How should titration and coulometry be selected?
A: Titration uses simpler apparatus and suits the routine water-content range; coulometry is more sensitive and suits precise low-content measurement. For low moisture with high accuracy demands (e.g., dew-point control), coulometry is preferred.
โ Q: How much do dry sampling conditions affect the result?
A: A great deal. Any moisture adsorption, leakage or condensation in the sampling and analysis system directly affects the result, so the dryness of lines, vessels, apparatus and reagents must be strictly controlled in accordance with the series.
โ Q: Is the method applicable to all natural gases?
A: It applies to natural gas matrices; for gases containing high concentrations of reducing components such as hydrogen sulphide, interference control and sample pre-treatment deserve particular attention, following the provisions of the standard.
โ Q: In what units should the result be expressed?
A: Results may be expressed as mass concentration (e.g., mg/mยณ), volume fraction or dew point; the series specifies the expression of results and report content โ keep the units consistent with the contract or technical agreement.