🧪 The BS EN ISO 06145 series is the British Standard implementation of EN ISO 6145, which is identical to ISO 6145. It specifies methods and equipment requirements for preparing calibration gas mixtures using dynamic volumetric methods, in which known flows of gas are continuously mixed and diluted. It is used for calibrating and validating analytical instruments in environmental monitoring, industrial gas analysis, vehicle emissions and natural gas analysis.
💡 The advantage of the dynamic volumetric method is that it continuously and stably produces a calibration gas at the desired concentration, which can be calculated precisely from the flow ratio, making it especially suitable for low concentrations, reactive components and large-batch calibration needs.
1. Series Overview 📌
Part 1 gives the general calibration methods; Part 10 specifies the permeation method; Part 11 specifies electrochemical generation; the remaining parts (Parts 4 to 9) specify the apparatus and procedures for different dilution/mixing principles (for example mass flow ratio, critical flow orifice, capillary, volumetric). The UK committee is PTI/15 (Natural gas and gas analysis), mirroring ISO/TC 158 “Analysis of gases”.
⚠️ Parts 4 to 9 of this series are scanned versions; the related descriptions here are compiled from public domain knowledge of the standard’s field, and the authoritative text is the official BSI published version. Some older editions (for example Part 1 and Part 10 of 2008) have been withdrawn.
2. Series Structure and Technical Points 🧩
The series contains 9 standard documents, as shown below.
| Standard No. | Edition | Subject (English title) |
| BS EN ISO 06145-1 | 2008 | Part 1: Methods of calibration |
| BS EN ISO 06145-10 | 2008 | Part 10: Permeation method |
| BS EN ISO 06145-11 | 2008 | Part 11: Electrochemical generation |
| BS EN ISO 06145-4 | 2008 | Part 4: Dynamic dilution apparatus (scanned) |
| BS EN ISO 06145-5 | 2010 | Part 5: Dynamic dilution apparatus (scanned) |
| BS EN ISO 06145-6 | 2008 | Part 6: Dynamic dilution apparatus (scanned) |
| BS EN ISO 06145-7 | 2010 | Part 7: Dynamic dilution apparatus (scanned) |
| BS EN ISO 06145-8 | 2008 | Part 8: Dynamic dilution apparatus (scanned) |
| BS EN ISO 06145-9 | 2010 | Part 9: Dynamic dilution apparatus (scanned) |
BS EN ISO 06145-1:2008 is identical to ISO 6145-1:2003 (prepared by ISO/TC 158 “Analysis of gases”), with ICS 71.040.40.
3. Key Technical Requirements 🔧
3.1 General calibration methods (Part 1)
Part 1 specifies the general requirements for preparing calibration gas mixtures by dynamic volumetric methods, including component selection, carrier gas purity, flow control and metering, mixing homogeneity, concentration calculation and uncertainty evaluation.
3.2 Permeation method (Part 10)
Part 10 specifies the method using a permeation tube to release standard gas at a stable rate at constant temperature, then diluting with carrier gas to the required concentration, suitable for trace-level organic/inorganic gas components.
3.3 Electrochemical generation (Part 11)
Part 11 specifies the apparatus and procedure for generating a target gas precisely by electrochemical reaction (electrolysis) — for example hydrogen or chlorine — and diluting it to the required concentration, suitable for high-purity, precisely metered gas generation.
🚨 The accuracy of flow meters and permeation tube temperature control directly determines the calibration gas concentration. Flow meters should be calibrated, the permeation temperature stabilised and carrier purity verified before use, otherwise systematic errors occur.
4. Practical Engineering Application 🛠️
- 🔍 Instrument calibration: generate multi-concentration calibration gas with the dynamic apparatus and calibrate linearity, repeatability and response time.
- 🔧 Low-concentration preparation: prefer permeation or electrochemical generation for trace-level standards to avoid adsorption losses of static cylinders.
- 📊 Metrological traceability: dynamic concentration is computed from flow ratios, so flow meters must be traceable to national/international standards.
- 📋 Uncertainty: evaluate concentration uncertainty per Part 1 and report it in the calibration certificate.
5. Frequently Asked Questions ❓
❓ Q: What is the difference between dynamic volumetric and static cylinder gas?
A: The dynamic method mixes continuously, so concentrations are flexible and free of adsorption/leakage losses, suiting multiple concentrations and trace components; static cylinders have fixed concentration and are stored after preparation, suiting batch use, but trace components tend to adsorb on cylinder walls.
❓ Q: What gas is the permeation method suitable for?
A: It suits trace gas components released stably by a permeation tube (for example SO₂, NO₂, H₂S), producing a stable permeation rate at constant temperature and then diluting.
❓ Q: Why is uncertainty evaluation needed?
A: The concentration error of the calibration gas is transferred directly to the instrument calibration result. Evaluating and combining the uncertainty from flow, temperature and other factors per Part 1 gives a credible concentration range.
🔍 Q: Which industries does the series serve?
A: Environmental monitoring (online analysers), industrial gas analysis, vehicle emissions testing, natural gas analysis and laboratory instrument calibration and quality control.