🎙️ BS EN 61094 is the European Standard series on electroacoustics measurement microphones, whose technical content is identical to IEC 61094. Measurement microphones are the metrological “yardstick” of acoustics. This series specifies laboratory standard microphones, primary pressure and free-field reciprocity calibration, working standard microphones and electrostatic-actuator tests, forming the traceability chain from “national standard” to “day-to-day working microphones”: the accuracy of sound-level meters, environmental noise monitoring, building acoustics and audio-product testing all rest on this series. The UK participates through EPL/29 (Electroacoustics).
The IEC 61094 series was developed by IEC TC 29 (Electroacoustics), adopted by CENELEC as EN 61094 and published by BSI (for example BS EN 61094-1:2001 (superseding the 1995 edition), BS EN 61094-2:2009). The series is organised as “standard microphones — working standard microphones — associated test methods”: Part 1 defines laboratory standard microphones (LS1/LS2 classes); Parts 2 and 3 give pressure and free-field absolute reciprocity calibration; Part 6 gives the electrostatic-actuator test for rapid frequency-response determination. Note that since 1997 IEC added the 60000 prefix to old numbers (for example IEC 27-1 became IEC 60027-1), while this series has always used the 61094 number.
| Standard | Subject | Content Focus |
|---|---|---|
| BS EN 61094-1:2001 | Specifications for laboratory standard microphones | Dimensions, frequency response and electrical performance of LS class |
| BS EN 61094-2:2009 | Primary method for pressure calibration by reciprocity | Coupler pressure reciprocity absolute calibration |
| BS EN 61094-3:2016 | Primary method for free-field calibration by reciprocity | Reciprocity calibration under free-field conditions |
| BS EN 61094-4:1996 | Specifications for working standard microphones | WS class working standard microphones |
| BS EN 61094-5:2016 | Pressure calibration of working standard microphones | Calibration of WS microphones by comparison/reciprocity |
| BS EN 61094-6:2005 | Electrostatic actuators | Frequency response measurement with electrostatic actuators |
| BS EN 61094-8:2012 | Determination of free-field sensitivity | Method for determining free-field sensitivity |
Part 1 specifies the mechanical dimensions, nominal sensitivity, frequency response, polarisation voltage and equivalent volume of laboratory standard microphones (LS class), ensuring that standard microphones from different laboratories and manufacturers share consistent, traceable geometry and electrical characteristics.
Part 2 gives the “primary method”: three microphones are paired in a coupler, and using the reciprocity principle the electrical transfer impedance is measured under known acoustic impedance to determine the absolute pressure sensitivity — the core procedure for tracing sound-pressure quantity to the national standard.
Free-field calibration (Part 3) accounts for the scattering of the sound field by the microphone and applies to free-field (0° incidence) conditions; the electrostatic actuator (Part 6) drives the microphone diaphragm equivalently with an electrostatic membrane, allowing fast, stable frequency-response determination without an acoustic coupler, suitable for routine verification.